Model 6492 Unipolar Temporary Atrial Pacing Lead
Medtronic Streamline Temporary Pacing Lead Comparison Chart
|
POSITIONING |
UNITS |
ELECTRODES |
FIXATION |
WIRE |
Model # |
Description |
Leads/
box |
Lead
color |
Type &
Length |
Atrial
Pacing
possible? |
Type/
Design |
Ancillary
Fixation |
Wire
diameter
(mm) |
|
UNIPOLAR PACING LEADS
|
6491 |
Pediatric |
12 |
Yellow |
3 mm discrete |
Yes |
Coil |
No |
0.5 |
6492 |
Atrial |
12 |
Purple |
3 mm discrete |
Yes |
Coil |
No |
0.48 |
6494 |
Adult |
24
(12 packets of 2) |
1-Yellow
&
1-Purple |
Nondiscrete |
Yes |
Straight |
No |
0.5 |
6500 |
Standard Unipolar Pacing Lead |
12 |
Clear |
4 mm discrete
(8 mm surface area to improve sensing) |
Yes |
Coil |
Includes Silicone Fixation Disk |
0.7 |
|
BIPOLAR PACING LEADS
|
6495 |
Adult Bipolar Coaxial |
6 |
Blue |
2 x 4 mm discrete (Controlled spacing between electrodes) |
Yes |
Coil |
No |
0.7 |
| |
CHEST NEEDLE |
MYOCARDIAL NEEDLE |
USE TIME |
SHELF LIFE |
PACKAGING |
Model # |
Breakaway needle (Needle breaks to expose lead plug) |
Needle diameter (mm) |
Curved or straight |
Thickness (mm) |
Curve |
Max. time for pacing & sensing |
Years |
|
|
UNIPOLAR PACING LEADS
|
6491 |
|
0.9 |
Curved |
0.4 |
7 mm radius |
7 days |
4 |
Double sterile Tyvek packaging |
6492 |
|
1.0 |
Straight |
0.4 |
7 mm radius |
7 days |
4 |
Double sterile Tyvek packaging |
6494 |
|
1.0 |
Straight |
0.55 |
12 mm radius |
7 days |
4 |
Double sterile Tyvek packaging |
6500 |
|
1.0 |
Straight |
0.66 |
12 mm radius |
7 days |
4 |
Double sterile Tyvek packaging |
|
BIPOLAR PACING LEADS
|
6495 |
|
1.6 |
Straight |
0.44 |
12 mm radius |
7 days |
4 |
Double sterile Tyvek packaging |
*EPG=external pulse generator & PSA = Pacing system analyser (the 2 different kinds of pacing generators).
|
|