Overview:

Portable and Accurate Measurement Technology

The Nellcor™ N-85 portable pulse oximetry monitor with Microstream™ technology provides both accurate SpO2 and etCO2 values in a single, easy-to-use, handheld device.

Microcap™ handheld capnographs deliver an easy-to-use assessment of respiratory status, whether intubated or non-intubated, in any clinical setting.

Microcap™ Handheld Capnograph Features

Features:

Versatile design for fast, simple set up and use

Handheld monitors that are convenient for spot checks and continuous monitoring in a variety of settings, including EMS/ED, transport.

Specifications

Specifications:

Specification Overview

Nellcor™ N-85 Hand-Held Capnograph/Pulse Oximeter with OxiMax™ Technology
Display Range: 0-99 mmHg, 0-13.2 kPa and 0-13.0 Vol% at sea level
Sampling Flow Rate: 50 ± 7.5 mL/min
Warm-Up Time: 30 seconds typical, reaches steady-state
Accuracy: 0 to 38 mm Hg ±2 mm Hg
  39 to 99 mm Hg ±5% of reading
Interface: RS-232 interface at 9600 baud; full duplex serial interface via RJ-45 connector
Modes: Adult and Neonate
Alarms: Adjustable alarm limits for all parameters; audio level as defined in ISO 9703-2 standard
Display: Graphic LCD display (128 x 64 dots) with LED backlight (contrast and backlight control); two numeric fields, three digits each, using 7-segment LED; two-color LED alarm bar
Microcap™ Handheld Capnograph
Sampling Flow Rate: 50 ml/ min
Warm-up Time: 30 seconds typical
Accuracy: 0-38 mmHg ± 2 mmHg
  39-99 mmHg ± 2 mmHg ± (5% of expected reading in mmHg +[0.08 x (expected reading in mmHg –39mmHg)])
  Expected reading = (barometric pressure in mmHg) x (%CO2 (in Vol%)) x 0.97
  Equivalent values for kPa and Vol%
  Respiratory rate 0-150 breaths/min
Modes: Adult and Neonate
Alarms: Adjustable alarm limits for etCO2, FiCO2, and respiratory rate
Display: Graphic LCD (128 x 64 dots) with LED backlight, (contrast and backlight control)
  Bright, seven-segment LED for numeric display of etCO2 and respiratory rate
  Two color alarm bar
Interface: Nurse Call RS232 interface between monitor and remote systems
  Bernoulli MSM Nurse Call System easily wired via stereo phone jack
  Philips®* Patient Monitor Interface: Microcap® only - one parameter (v3.15 and higher)