Medtronic CRM Product Performance Report

7274 Marquis DR

    Product Characteristics
US Market Release Mar-02   NBD Code VVED
Registered US Implants 48,000   Serial Number Prefix PKC
Estimated Active US Implants 15,000   Max Delivered Energy 30 J
Normal Battery Depletions (US) 1,495   Estimated Longevity  
Malfunctions (US) 150      
Therapy Function Not Compromised 75
     
Battery 5   (3 malfunctions related to advisory)  
Electrical Component 23      
Possible Early Battery Depletion 47      
Therapy Function Compromised 75      
Battery 52   (42 malfunctions related to advisory)  
Electrical Component 23      
Advisories 2005 Potential Premature Battery
Depletion Due to Battery Short
Survival Estimate of Marquis DR ICD Model 7274