Medtronic CRM Product Performance Report

7275 GEM III DR

    Product Characteristics
US Market Release Nov-00   NBD Code VVED
Registered US Implants 20,000   Serial Number Prefix PJM
Estimated Active US Implants 2,000   Max Delivered Energy 30 J
Normal Battery Depletions (US) 3,471   Estimated Longevity  
Malfunctions (US) 41      
Therapy Function Not Compromised 30
     
Battery 1      
Electrical Component 11      
Software/Firmware 1      
Possible Early Battery Depletion 17      
Therapy Function Compromised 11      
Battery 2      
Electrical Component 8      
Electrical Interconnect 1      
Performance Note: Performance note on ICD Battery Discharge Behavior
Survival Estimate of Gem III DR Model 7275