Medtronic CRM Product Performance Report

6721, 6921 Epicardial Patch

    Product Characteristics
US Market Release Feb-93   Serial Number Prefix TBH, TBG, TBB, TAD, TAC, or TAB
Registered US Implants 8,300   Type and/or Fixation Epicardial Defib Patch, Suture
Estimated Active US Implants 1,300   Polarity Defib Electrode only
Advisories None   Steroid No

 

System Longevity Study Results  
US Returned Product Analysis
Conductor Fracture 68
Crimp/Weld/Bond 1
Insulation Breach 10
Extrinsic 5
Other 1
Number of Leads Enrolled in Study 407
Cumulative Months of Follow-Up 19,854
Number of Leads Active in Study 12
   
Qualifying Complications 52 Total  
Conductor Fracture 21  
Failure to Capture 8  
Impedance Out of Range 4  
Insulation (not further defined) 3  
Oversensing 16  

 

Survival Estimate of Epicardial Patch 6721, 6921