Home > Leads > Defibrillation
6942 Sprint |
|
|
Product Characteristics |
| US Market Release |
Jul-97 |
|
Serial Number Prefix |
TCB |
| Registered US Implants |
17,700 |
|
Type and/or Fixation |
Transvenous, Vent, Defib and
Pace/Sense, Tines |
| Estimated Active US Implants |
7,400 |
|
Polarity |
Integrated Bipolar/Two Coils |
| Advisories |
None |
|
Steroid |
Yes |
| System Longevity Study Results |
|
| US Returned Product Analysis |
| Implant Damage |
31 |
| Electrical Malfunction |
37 |
| Other |
5 |
|
| Number of Leads Enrolled in Study |
351 |
| Cumulative Months of Follow-Up |
15,650 |
| Number of Leads Active in Study |
57 |
| |
|
| Qualifying Complications |
7 Total |
|
| Conductor Fracture |
1 |
|
| Failure to Sense |
1 |
|
| Lead Dislodgement |
1 |
|
| Oversensing |
3 |
|
| Unspecified Clinical Failure |
1 |
|
| |
|
|