Medtronic CRM Product Performance Report

4968 CapSure Epi

    Product Characteristics
US Market Release Sep-99   Serial Number Prefix LEN
Registered US Implants 16,000   Type and/or Fixation Epicardial Suture-On V or A
Estimated Active US Implants 10,000   Polarity Bipolar
Advisories None   Steroid Yes

 

System Longevity Study Results  
US Returned Product Analysis
Implant Damage 2
Electrical Malfunction 11
Other 0
Number of Leads Enrolled in Study 543
Cumulative Months of Follow-Up 22,604
Number of Leads Active in Study 326
     
Qualifying Complications 34 Total  
Conductor Fracture 7  
Failure to Capture 14  
Failure to Sense 3  
Impedance Out of Range 4  
Insulation (not further defined) 2  
Oversensing 4  

 

Survival Estimate of 4968 CapSure Epi Leads