Home > Leads > Epi/Myocardial
6917, 6917A Tenax |
|
|
Product Characteristics |
| US Market Release |
Jun-73 |
|
Serial Number Prefix |
WV or WC |
| Registered US Implants |
180,100 |
|
Type and/or Fixation |
Myocardial Screw-in Vent. |
| Estimated Active US Implants |
5,000 |
|
Polarity |
Unipolar |
| Advisories |
None |
|
Steroid |
No |
| System Longevity Study Results |
|
| US Returned Product Analysis |
| Implant Damage |
115 |
| Electrical Malfunction |
42 |
| Other |
1 |
|
| Number of Leads Enrolled in Study |
985 |
| Cumulative Months of Follow-Up |
47,481 |
| Number of Leads Active in Study |
4 |
| |
|
|
| Qualifying Complications |
69 Total |
|
| Conductor Fracture |
6 |
|
| Extra Cardiac Stimulation |
1 |
|
| Failure to Capture |
30 |
|
| Failure to Sense |
11 |
|
| Impedance Out of Range |
2 |
|
| Insulation (MIO) |
1 |
|
| Oversensing |
18 |
|
| |
|
|