4074 CapSure Sense |
Product Characteristics | |||
|---|---|---|---|---|
| US Market Release | Jun-02 | Serial Number Prefix | BBD | |
| Registered US Implants | 63,000 | Type and/or Fixation | Transvenous, Vent., Tines | |
| Estimated Active US Implants | 45,200 | Polarity | Bipolar | |
| Advisories | None | Steroid | Yes | |
| Ventricular Placement | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| System Longevity Study Results |
|
|||||||||
| Number of Leads Enrolled in Study | 622 | |||||||||
| Cumulative Months of Follow-Up | 24,361 | |||||||||
| Number of Leads Active in Study | 516 | |||||||||
| Qualifying Complications | 5 Total | |||||||||
| Failure to Capture | 1 | |||||||||
| Failure to Sense | 1 | |||||||||
| Impedance Out of Range | 1 | |||||||||
| Lead Dislodgement | 2 | |||||||||