Medtronic CRM Product Performance Report

4081 Target Tip

    Product Characteristics
US Market Release Jul-89   Serial Number Prefix LAC
Registered US Implants 3,900   Type and/or Fixation Transvenous, Vent., Tines
Estimated Active US Implants 800   Polarity Unipolar
Advisories None   Steroid No

 

Ventricular Placement    
System Longevity Study Results  
US Returned Product Analysis
Implant Damage 4
Electrical Malfunction 5
Other 0
Number of Leads Enrolled in Study 260
Cumulative Months of Follow-Up 9,940
Number of Leads Active in Study 9
     
Qualifying Complications 3 Total  
Conductor Fracture 1  
Failure to Sense 2  
     

Survival Estimate of 4081 Target Tip Ventricular Placement Leads