5023, 5023M CapSure SP |
Product Characteristics | |||
|---|---|---|---|---|
| US Market Release | Nov-88 | Serial Number Prefix | SX or LAS | |
| Registered US Implants | 9,900 | Type and/or Fixation | Transvenous, Vent., Tines | |
| Estimated Active US Implants | 2,300 | Polarity | Unipolar | |
| Advisories | None | Steroid | Yes | |
| Ventricular Placement | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| System Longevity Study Results |
|
|||||||||
| Number of Leads Enrolled in Study | 1,353 | |||||||||
| Cumulative Months of Follow-Up | 70,587 | |||||||||
| Number of Leads Active in Study | 470 | |||||||||
| Qualifying Complications | 15 Total | |||||||||
| Conductor Fracture | 2 | |||||||||
| Extra Cardiac Stimulation | 4 | |||||||||
| Failure to Capture | 8 | |||||||||
| Impedance Out of Range | 1 | |||||||||