Medtronic CRM Product Performance Report

5024, 5024M CapSure SP

    Product Characteristics
US Market Release Mar-90   Serial Number Prefix SY or LAT
Registered US Implants 201,500   Type and/or Fixation Transvenous, Vent., Tines
Estimated Active US Implants 56,200   Polarity Bipolar
Advisories None   Steroid Yes

 

Ventricular Placement    
System Longevity Study Results  
US Returned Product Analysis
Implant Damage 723
Electrical Malfunction 115
Other 29
Number of Leads Enrolled in Study 8,140
Cumulative Months of Follow-Up 431,840
Number of Leads Active in Study 594
     
Qualifying Complications 48 Total  
Conductor Fracture 3  
Extra Cardiac Stimulation 2  
Failure to Capture 25  
Failure to Sense 2  
Impedance Out of Range 3  
Insulation (ESC) 1  
Insulation (not further defined) 5  
Lead Dislodgement 5  
Oversensing 1  
Unspecified Clinical Failure 1  

Survival Estimate of 5024, 5024M CapSure SP Ventricular Placement Leads