5034 CapSure Z |
Product Characteristics | |||
|---|---|---|---|---|
| US Market Release | Feb-96 | Serial Number Prefix | LDF | |
| Registered US Implants | 56,300 | Type and/or Fixation | Transvenous, Vent., Tines | |
| Estimated Active US Implants | 17,000 | Polarity | Bipolar | |
| Advisories | None | Steroid | Yes | |
| Ventricular Placement | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| System Longevity Study Results |
|
|||||||||
| Number of Leads Enrolled in Study | 1,594 | |||||||||
| Cumulative Months of Follow-Up | 82,761 | |||||||||
| Number of Leads Active in Study | 228 | |||||||||
| Qualifying Complications | 13 Total | |||||||||
| Conductor Fracture | 1 | |||||||||
| Failure to Capture | 9 | |||||||||
| Failure to Sense | 1 | |||||||||
| Impedance Out of Range | 1 | |||||||||
| Lead Dislodgement | 1 | |||||||||