Medtronic CRM Product Performance Report

5068 CapSureFix

    Product Characteristics
US Market Release Jan-97   Serial Number Prefix LDJ
Registered US Implants 103,100   Type and/or Fixation Transvenous, V or A, Screw-in
Estimated Active US Implants 40,600   Polarity Bipolar
Advisories None   Steroid Yes

 

Atrial Placement    
System Longevity Study Results  
US Returned Product Analysis
Implant Damage 455
Electrical Malfunction 75
Other 15
Number of Leads Enrolled in Study 968
Cumulative Months of Follow-Up 32,929
Number of Leads Active in Study 65
 
Qualifying Complications 6 Total  
Failure to Capture 2  
Impedance Out of Range 2  
Lead Dislodgement 1  
Oversensing 1  
     

Survival Estimate of 5068 CapSureFix Atrial Placement Leads

Ventricular Placement    
System Longevity Study Results    
Number of Leads Enrolled in Study 1,359
Cumulative Months of Follow-Up 37,307
Number of Leads Active in Study 119
     
Qualifying Complications 6 Total  
Conductor Fracture 1  
Failure to Capture 3  
Insulation (not further defined) 1  
Lead Dislodgement 1  
     
Survival Estimate of 5068 CapSureFix Ventricular Placement Leads