Medtronic CRM Product Performance Report

5524, 5524M CapSure SP

    Product Characteristics
US Market Release Mar-90   Serial Number Prefix XV or LAV
Registered US Implants 60,600   Type and/or Fixation Transvenous, Atrial-J, Tines
Estimated Active US Implants 21,000   Polarity Bipolar
Advisories None   Steroid Yes

 

Atrial Placement    
System Longevity Study Results  
US Returned Product Analysis
Implant Damage 67
Electrical Malfunction 25
Other 7
Number of Leads Enrolled in Study 4,445
Cumulative Months of Follow-Up 241,326
Number of Leads Active in Study 482
     
Qualifying Complications 38 Total  
Conductor Fracture 1  
Failure to Capture 22  
Failure to Sense 4  
Impedance Out of Range 1  
Insulation (not further defined) 2  
Lead Dislodgement 4  
Oversensing 4  
     

Survival Estimate of 5524, 5524M CapSure SP Atrial Placement Leads