|
RF Generators
| Atakr® II RF Ablation
System |
 |
| (60 Watt output) |
(Atakr®
II) The Atakr II RF Generator provides high power output with
the integral safety of closed-loop temperature control using the
embedded thermocouples in Medtronic RF ablation catheters.
Features:
- 60 Watt output
- Closed-loop temperature control
- Operates in power or temperature mode
- LEM feature provides tip to tissue contact
- Adjustable impedance parameters
- Adjustable power and temperature limits for safety
- Remote control electronic foot pedal
- Line powered
The System comes equipped with the following:
1 60 Watt RF Generator
1 Electronic foot pedal
1 Atakr to RF catheter cable
1 Dispersive electrode cable
1 Atakr to EGM cable
return to top
Atakr® II RF Ablation
System
Indications for use
The Atakr II ablation system is indicated for use for the following:
- Interruption of accessory atrioventricular (AV) conduction pathways
associated with tachycardia
- Treatment of AV nodal re-entrant tachycardia
- Creation of complete AV block in patients with a difficult to
control ventricular response to an atrial arrhythmia
Contraindications
The Atakr II ablation system is contraindicated for the following:
- Patients with active systemic infection
- The transeptal approach in patients with left atrial thrombus
or myxoma or interatrial baffle or patch
- The retrograde transaortic approach in patients with aortic
valve replacement
Warnings and precautions
System compatibility Only use Medtronic RF catheters
and accessories. The safety of use with other electrophysiology
catheters or accessories has not been assessed.
Long term risks The long-term effects of lesions
created by
RF ablation have not been established.
- Any long-term effects of lesions in proximity to the conduction
system or coronary vasculature are unknown.
- The risk/benefit in asymptomatic patients has not been studied.
AV node modification or septal accessory pathway ablation
Immediately terminate RF energy delivery if partial or complete
AV block is noted. Patients undergoing AV node modification or septal
accessory pathway ablation are at risk for complete AV block. Closely
monitor AV conduction during RF energy delivery.
Left-sided ablation Patients undergoing left-sided
ablation procedures should be closely monitored during the post-ablation
period for clinical manifestations of infarction, thromboembolus
or stroke.
Left-sided catheter insertions Anticoagulate patients
undergoing left-sided catheter insertions.
High ablation temperature High ablation electrode
temperatures, often caused by coagulum formation on the ablation
electrode, may lead to charring of the cardiac tissue, or a vaporization
of interstitial and intracellular fluid. To reduce the possibility
of hazards associated with high temperatures:
- Use an RF catheter that features a thermocouple and adjust the
temperature setpoint/limit to the maximum acceptable measured
temperature (± 5C) for the ablation electrode. See Setting
the control mode, parameters, and options in the manual
for further information on adjusting the temperature setpoint/limit.
- Remove the RF catheter and examine the ablation electrode for
coagulum. Remove any coagulum found.
Handling RF catheters
- Use fluoroscopy during catheter manipulation and placement.
RF power application within a coronary artery has been associated
with myocardial infarction and death.
- Do not touch the ablation electrode of the RF catheter and the
dispersive electrode at the same time while operating the RF power
generator. Operator injury may occur.
Sudden impedance rise Discontinue application of
RF energy if a
sudden impedance rise is observed. Remove the RF catheter and
examine the ablation electrode for coagulum. Remove any
coagulum found.
Non-thermocouple (NTC) RF catheters NTC RF catheters
do not
provide temperature measurement. The RF power generator will operate
only in power control mode when using NTC RF catheters:
- When using an NTC RF catheter, use an RF catheter interface cable
designed for using NTC RF catheters with the Atakr II ablation
system. Follow the directions provided in the RF catheter interface
cable product literature when using the RF power generator with
these NTC RF catheters.
- Do not use an RF interface cable designed for NTC RF catheters
with
an RF catheter that features a thermocouple.
X-ray and fluoroscopic exposure Minimize x-ray exposure.
Significant x-ray exposure can result in acute radiation injury
as well as
dose-related risk for somatic and genetic effects. Take all appropriate
measures to minimize x-ray exposure to both patients and clinical
staff.
Electrosurgery near flammable materials The risk
of igniting
flammable gases or other materials is inherent in electrosurgery
and
cannot be eliminated by device design. Take precautions to restrict
flammable materials and substances from the electrosurgical site.
Flammable materials may come in the following forms:
- Anesthetics or skin preparation agents
- Natural chemicals produced within body cavities
- Chemicals originating in surgical drapes or other materials
Incidental patient burns Avoid unintentional burns on
the patient
during RF energy delivery:
- Minimize the distance between the dispersive electrode and the
operating field.
- Minimize skin-to-skin contact between parts of the patient's body
by
covering these areas with dry gauze.
- Place monitoring electrodes as far as possible from surgical
electrodes to minimize burns at the site of the monitoring electrodes.
Using RF energy near implanted devices Implanted devices,
such
as pacemakers and implantable cardioverter-defibrillators (ICDs),
may be
adversely affected by RF energy:
- Keep external sources of pacing and defibrillation available
during ablation.
- Deactivate ICD detection during the delivery of RF energy.
- Exercise extreme caution when delivering ablation energy in close
proximity to implanted leads.
- Perform complete implantable device testing before and
after ablation.
DC energy Do not deliver DC energy through the RF
power generator.
No testing has been performed on the RF power generator regarding
the
delivery of DC energy through it.
Fluid incursion The RF power generator may not function
correctly if
the electronic circuitry or the connectors are wet:
- Do not allow any fluid or moisture into the RF power generator
or any
connector or cables.
- Do not hang fluids above the RF power generator.
- Do not immerse the cables into fluids.
Electrical shock Do not touch the patient while touching
the outer housing or connections on the Atakr II, to avoid the risk
of electrical shock to the patient. Do not allow the patient to
come into contact with grounded metal surfaces during RF energy
delivery.
Line-powered equipment An inserted RF catheter or
an RF catheter
connected to an RF catheter cable constitutes a direct, low-resistance
current pathway to the myocardium. Due to the danger of
tachyarrhythmias resulting from alternating current leakage, extreme
caution must be taken to properly ground all line-powered equipment
used on or in the vicinity of the patient.
Monitoring systems Monitoring systems incorporating
high-frequency, current-limiting devices are recommended. Monitoring
electrodes should be placed as far as possible from surgical electrodes.
Needle monitoring systems are not recommended.
Information technology equipment All information
technology
equipment that is attached to the Atakr II ablation system must
be
approved by a third party to the requirements of UL 1950 or EN 60950.
Stimulation equipment Certain stimulation equipment
alters the
RF energy waveform, which may lead to rare occasions of atrial or
ventricular fibrillation. To eliminate the influence of stimulation
equipment
during ablation, use the options menu to disconnect the electrogram
(EGM) device.
Electromagnetic interference (EMI) This device is
a non-ionizing
emitter of radiation and may cause EMI with other devices. In order
to
minimize EMI, the generator should be at least 6 in. away from any
other
electronic device.
- If EMI is apparent during the application of RF energy, EMI may
be
reduced by repositioning the RF power generator or other equipment.
- If EMI occurs in electrogram or ECG recording equipment during
RF energy delivery after re-adjusting the ablation electrode, use
the
options menu to disconnect the EGM device. If the EMI persists,
unplug the ablation EGM electrode.
RF power generator failure Failure of the Atakr II
RF power generator
could result in an unintended increase of output power. However,
RF energy delivery stops if the operator selected or manufacturer
programmed impedance, power, or temperature limits are exceeded.
RF power generator storage temperature If the RF
power generator has been stored at temperatures greater than 30°C
or less than 15°C, allow the RF power generator to reach room
temperature by waiting at least 30 minutes before use. The RF power
generator may not function correctly if the electronic circuitry
is too cold or too hot.
Adverse events
The potential adverse events related to the use of ablation systems
include, but are not limited to, the following:
- Death
- Cardiac tamponade
- Cerebrovascular accident
- Myocardial infarction
- Coronary sinus perforation
- Bacterial endocarditis
- Ventricular fibrillation
- Coronary artery spasm
- Valvular insufficiency
- Femoral artery laceration
- Thrombus/embolic event
- Bowel obstruction
- Brachial plexus injury
- Pneumonia
- Inferior vena cava clot
- Deep vein thrombosis
- Pericardial effusion
- Pericarditis
- Pleural effusion
- Respiratory depression
- Atrial lead dislodgment
- Infected IV site
- Hypotension
- Bradycardia
- Vasovagal reaction
- Thrombophlebitis
- Temperature elevation
- Puncture site hematoma
- Phlebothrombosis
- High CPK
- Pulmonary embolism
- Thromboembolism
- Nerve or blood vessel injury
- Phrenic nerve injury
- Transient ischemic attack
- Infection
- Unintended AV or bundle
branch block requiring
pacemaker implantation
return to top
|