Sensia
Pacing System
sensia-intro

Sensia Pacing System

The Sensia® Pacing System provides proven pacing with expanded options.

Overview

Pacing Redefined – The New Standard

The new standard of care in pacing is to reduce unnecessary right ventricular pacing to as close to zero as possible. Mounting evidence suggests that right ventricular pacing is associated with a variety of detrimental effects, including risk of heart failure hospitalization and atrial fibrillation.1

Important Safety Information

Changes in a patient's disease and/or medications may alter the efficacy of a device's programmed parameters or related features.

The Sensia pacing system combines physiologic pacing with automaticity.

Risk of Heart Failure Hospitalization (HFH)

Risk of Heart Failure Hospitalization (HFH)1

Search AV™+ Promotes Intrinsic Conduction

  • Search AV+ (SAV+) automatically and continuously searches for natural intrinsic conduction
  • SAV+ reduces unnecessary right ventricular pacing to less than 20%*2

* Median number

References

  1. Sweeney M, Hellkamp A, Ellenbogen K, et al. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation. 2003;107:2932-2937.
  2. Melzer C, Sowelam S, Sheldon T, et al. Reduction of right ventricular pacing in patients with sinus node dysfunction using an enhanced search AV algorithm. Pacing Clin Electrophysiol. 2005;28(6):521-527.

How the Sensia Pacing System Works

Completely Automatic – Simple to Use

Atrial Capture Management and Ventricular Capture Management (ACM and VCM ):

  • Provide complete long-term threshold management automatically
  • Ensure pacing outputs remain at safe levels1
  • Adapts programmed outputs to actively manage battery current drain
TherapyGuide simplifies programming

TherapyGuide™ simplifies programming

TherapyGuide™ – Tailored, Clinically-Oriented Programming Using Physician Selected Parameters

  • Simplifies and tailors programming
  • Provides suggested device settings based on physician-selected patient conditions
Quick Look II Image 

Detail - Quick Look II image

Quick Look™ II – All Key Information on One Screen

  • Simplifies clinic follow-up with all key information on one screen
  • Provides concise overview of patient, device, and lead status
  • Routine follow-up measurements are performed automatically
  • All measurements are automatic and displayed upon initial interrogation

Medtronic CareLink Network for Pacemakers

Medtronic CareLink® Network, the leading Internet-based remote monitoring service, is now available for pacemakers.

Reference

  1. Sperzel J, Milasinovic G, Smith TW, et al. Automatic measurement of atrial pacing thresholds in dual-chamber pacemakers: Clinical experience with atrial capture management. Heart Rhythm. 2005;2(11):1203-1210.

Specifications

Sensia S

Mechanical
Model
ID
Size (HxWxD mm)M (g)V (cc)ConnectorRadiopaque
SES0140.2x42.9x7.521.59.7IS-1 BI or UNIPWB or PWL
Battery
TypeLithium-Iodine
Voltage2.8 V
Average Projected Capacity.91 Ah
Longevity8.4 years*

* 60 ppm, 100% pacing, 2.0V, 0.4 ms pulse width, 1,000 Ω pacing impedance.

Sensia D with SAV+

Mechanical
Model
ID
Size (HxWxD mm)M (g)V (cc)ConnectorRadiopaque
SED0144.7x47.9x7.527.112.1IS-1 BI or UNIPWL or PWB
Battery
TypeLithium-Iodine
Voltage2.8 V
Average Projected Capacity
  Model SED011.3 Ah
Longevity
  Model SED0111.3 years* 12.0 years**

* DDD, 60 ppm, 100% pacing, ventricular 2.0V† atrial 1.5V,† 0.4 ms pulse width, 1,000 Ω pacing impedance.
** SAV+ON 50% pacing

Sensia SR

Mechanical
Model
ID
Size (HxWxD mm)M (g)V (cc)ConnectorRadiopaque ID
SESR0140.2x42.9x7.521.59.7IS-1 BI or UNIPWL, PWB
Battery
TypeLithium-Iodine
Voltage2.8 V
Average Projected Capacity.91 Ah
Longevity8.4 years*

* SSR or SSI, 60 ppm, 100% pacing, 2.0V, 0.4 ms pulse width, 1,000 Ω pacing impedance.

Sensia DR with SAV+

Mechanical
Model
ID
Size (HxWxD mm)M (g)V (cc)ConnectorRadiopaque ID
SEDR0144.7x47.9x7.527.112.1IS-1 BI or UNIPWL, PWB
SEDRL145.4x52.3x7.531.113.1IS-1 BI or UNIPWL, PWB
(longevity)
Battery
TypeLithium-Iodine
Voltage2.8 V
Average Projected Capacity
  Model SEDR011.3 Ah
  Model SEDRL11.6 Ah
Longevitywith SAV+ offwith SAV+ on
  Model SEDR018.9 years*9.4 years**

* DDD, 60 ppm, 100% pacing, ventricular 2.0V† atrial 1.5V,† 0.4 ms pulse width, 1,000 Ω pacing impedance.
** SAV+ON 50% pacing

Last updated: 22 Sep 2010