NC Sprinter RX Noncompliant Balloon Dilatation Catheter
The NC Sprinter® RX Noncompliant Rapid Exchange Balloon Dilatation Catheter combines tried and true Sprinter technology with the high-pressure capability and controlled expansion of a noncompliant balloon.
How the NC Sprinter RX Works
Stent implantation often follows balloon angioplasty. Once the stent is implanted, post stent deployment as assessed by IVUS or fluoroscopy may be completed to ensure adequate stent expansion and apposition against the arterial wall.1

Features
The NC Sprinter RX noncompliant rapid exchange balloon dilatation catheter combines tried and true Sprinter technology with the high-pressure capacity and controlled expansion of a noncompliant balloon.2


Engineered for Exact Sizing to Enhance Stent Apposition

References
- Brodie BR. Adjunctive balloon postdilatation after stent deployment: is it still necessary with drug-eluting stents? J Interv Cardiol. 2006;19:43-50.
- NC Sprinter balloon models 2.25 mm to 5.0 mm are indicated for the post-delivery expansion of balloon expandable stents.
- Test results based on distal balloon shoulder. Test data on file at Medtronic, Inc. Bench test results may not be indicative of clinical performance.
- Test data on file at Medtronic, Inc. Bench test results may not be indicative of clinical performance.
Specifications
| Catheter length | 138 cm |
| Lesion entry profile (in.) | 0.016 |
Tip length (mm)
|
3.5 mm 4.5 mm |
| Balloon material | Soft Fulcrum Plus |
| Coating | Selective DuraTrac |
| Gold-swaged marker bands | Double |
| Nominal pressure | 10 atm |
| Rated burst pressure | 18 atm |
MiniWrap folds
|
3 folds 5 folds |
Shaft dimensions for 2.0 – 3.75 mm
|
1.9 F 2.4 F / 2.6 F |
Shaft dimensions for 4.00 – 5.00 mm
|
1.9 F 3.0 F |
| MGCID (in.) |
|
|
Diameters (mm) |
2.00 – 5.00 |
| Lengths (mm) | 6, 9, 12, 15, 21, 27 |
Crossing Profile
| Balloon Diameter (mm) | Crossing Profile OD (in) |
|---|---|
| 2.00 | 0.024 |
| 2.25 | 0.024 |
| 2.50 | 0.025 |
| 2.75 | 0.025 |
| 3.00 | 0.026 |
| 3.25 | 0.027 |
| 3.50 | 0.027 |
| 3.75 | 0.028 |
| 4.00 | 0.029 |
| 4.50 | 0.030 |
| 5.00 | 0.032 |
Compliance Table
| Pressure (atm) | Balloon Diameter (mm) | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| 2.00 | 2.25 | 2.50 | 2.75 | 3.00 | 3.25 | 3.50 | 3.75 | 4.00 | 4.50 | 5.00 | |
| 8 | 1.98 | 2.17 | 2.45 | 2.70 | 2.97 | 3.09 | 3.36 | 3.62 | 3.87 | 4.27 | 4.75 |
| 9 | 2.00 | 2.20 | 2.49 | 2.74 | 3.01 | 3.15 | 3.42 | 3.68 | 3.93 | 4.34 | 4.84 |
| 10 | 2.02 | 2.23 | 2.51 | 2.77 | 3.04 | 3.18 | 3.45 | 3.72 | 3.97 | 4.39 | 4.91 |
| 11 | 2.03 | 2.24 | 2.53 | 2.80 | 3.07 | 3.22 | 3.49 | 3.76 | 4.01 | 4.44 | 4.97 |
| 12 | 2.05 | 2.26 | 2.55 | 2.82 | 3.10 | 3.25 | 3.52 | 3.79 | 4.05 | 4.48 | 5.01 |
| 13 | 2.07 | 2.28 | 2.57 | 2.84 | 3.13 | 3.28 | 3.55 | 3.82 | 4.08 | 4.51 | 5.06 |
| 14 | 2.08 | 2.30 | 2.58 | 2.87 | 3.15 | 3.31 | 3.58 | 3.85 | 4.11 | 4.55 | 5.10 |
| 15 | 2.10 | 2.32 | 2.60 | 2.89 | 3.18 | 3.33 | 3.61 | 3.88 | 4.15 | 4.58 | 5.14 |
| 16 | 2.12 | 2.33 | 2.62 | 2.91 | 3.20 | 3.36 | 3.63 | 3.90 | 4.18 | 4.61 | 5.18 |
| 17 | 2.13 | 2.36 | 2.63 | 2.93 | 3.23 | 3.39 | 3.65 | 3.93 | 4.22 | 4.65 | 5.22 |
| 18 | 2.15 | 2.38 | 2.65 | 2.95 | 3.26 | 3.41 | 3.68 | 3.96 | 4.25 | 4.68 | 5.26 |
| 19 | 2.17 | 2.40 | 2.66 | 2.98 | 3.29 | 3.44 | 3.71 | 3.99 | 4.29 | 4.72 | 5.30 |
| 20 | 2.19 | 2.42 | 2.68 | 3.00 | - | 3.47 | 3.73 | 4.02 | - | 4.76 | - |
| 21 | 2.21 | 2.44 | 2.70 | 3.03 | - | 3.50 | 3.77 | 4.05 | - | 4.80 | - |
| 22 | 2.23 | 2.46 | 2.72 | 3.05 | - | 3.53 | - | 4.08 | - | 4.84 | - |
| Nominal pressure | |||||||||||
| Rated burst pressure | |||||||||||
Test data on file with Medtronic, Inc.
* Rated burst pressure: the maximum pressure to which a balloon is designed to be inflated (95% confidence that 99.9% will not fail below rated burst pressure). DO NOT EXCEED.
Nominal pressure: the pressure at which the balloon reaches its labeled diameter.