Sprinter Legend RX Semicompliant Balloon Dilatation Catheter
With innovative crossing technologies and 57 sizes, the Sprinter® Legend RX Semicompliant Balloon Dilatation Catheter provides you with the power to cross.
How the Sprinter Legend RX Catheter Works
Percutaneous Coronary Intervention (PCI), also known as coronary angioplasty, is a therapeutic procedure to treat the narrowed or stenotic section of the coronary artery of the heart due to coronary atherosclerosis. In balloon catheter angioplasty, tightly folded and narrow balloons are passed into the narrowed artery and inflated to a fixed size using liquid. This procedure reduces the narrowing of the artery by pushing the plaque (or blockage) to the sides of the artery.1

Stent implantation often follows balloon angioplasty. Once the stent is implanted, post stent deployment as assessed by IVUS or fluoroscopy may be completed to ensure adequate stent expansion and apposition against the arterial wall.
Designed to meet a spectrum of clinical challenges, Sprinter Legend RX offers:
- 1.50 – 4.00 mm low profile balloons: The best options for treating a comprehensive spectrum of lesions
- Micro-brite marker bands: Enhanced cross and track with uncompromised radiopacity
- Broad product offering: 57 sizes, 10 diameters, including ¼ sizes and 7 lengths
And Sprinter Legend RX features Medtronic's renowned balloon technologies.
![]() |
FasTrac TipWith 0.41-mm lesion entry profile for superior crossability |
![]() |
MiniWrap FoldingProprietary process gives 1.50-4.00 mm balloons low-profile shoulder and tight wrap/rewrap |
![]() |
Dura-Trac Selective CoatingFor excellent lubricity during tracking and position retention during inflation |
Sprinter Legend 1.5 mm

Sprinter Legend 3.0 mm
Sprinter Legend RX Animation (3:23, 21.3 MB)
This video gives an overview of the Sprinter Legend RX catheter's high performance features.
Reference
- Source from Medical Review: Caroline S. Rhoads, MD–Internal Medicine and Stephen Fort, MD, MCRP, FRCPC–Interventional Cardiology ©1995-2005, Healthwise.
Specifications
| Specifications | ||
|---|---|---|
| Catheter length | 142 cm | |
| Distal shaft length | 21 cm | |
| Lesion entry profile (in.) | 0.016 | |
| Tip length (mm) | 2.50 | |
| 1.50 mm crossing profile (in.) | 0.021 | |
| Balloon material | Fulcrum | |
| Coating | Selective DuraTrac | |
| Platinum-iridium marker bands | ||
| Single Double | |
| Nominal pressure | ||
| 6 atm 8 atm | |
| Rated burst pressure | ||
| 12 atm 14 atm | |
| MiniWrap folds | ||
| 2 folds 3 folds 5 folds | |
| Shaft dimensions for 1.50 – 3.50 mm | ||
| 1.9 F 2.4 F / 2.6 F | |
| Shaft dimensions for 3.75 – 4.00 mm | ||
| 1.9 F 2.5 F / 2.7 F | |
| MGCID (in.) | 0.056 | |
| Diameters (mm) | 1.50 – 4.00 | |
| Lengths (mm) | 6, 10, 12, 15, 20, 25, 30 | |
| Crossing Profile | |
|---|---|
| Balloon diameter (mm) | Crossing profile OD (in) |
| 1.50 | 0.021 |
| 2.00 | 0.022 |
| 2.25 | 0.022 |
| 2.50 | 0.022 |
| 2.75 | 0.023 |
| 3.00 | 0.023 |
| 3.25 | 0.023 |
| 3.50 | 0.023 |
| 3.75 | 0.023 |
| 4.00 | 0.024 |
Compliance
| Pressure (atm) | 1.5 | 2.00 | 2.25 | 2.50 | 2.75 | 3.00 | 3.25 | 3.50 | 3.75 | 4.00 |
|---|---|---|---|---|---|---|---|---|---|---|
| 4 | 1.36 | — | — | — | — | — | — | — | — | — |
| 5 | 1.41 | — | — | — | — | — | — | — | — | — |
| 6 | 1.47 | 1.91 | 2.23 | 2.34 | 2.60 | 2.81 | 3.14 | 3.32 | 3.61 | 3.85 |
| 7 | 1.53 | 1.95 | 2.29 | 2.41 | 2.67 | 2.89 | 3.22 | 3.42 | 3.70 | 3.96 |
| 8 | 1.58 | 2.00 | 2.33 | 2.47 | 2.74 | 2.97 | 3.29 | 3.50 | 3.79 | 4.05 |
| 9 | 1.61 | 2.03 | 2.38 | 2.53 | 2.79 | 3.04 | 3.35 | 3.57 | 3.86 | 4.12 |
| 10 | 1.64 | 2.06 | 2.41 | 2.57 | 2.84 | 3.09 | 3.40 | 3.63 | 3.93 | 4.18 |
| 11 | 1.67 | 2.09 | 2.44 | 2.61 | 2.89 | 3.14 | 3.44 | 3.68 | 3.98 | 4.23 |
| 12 | 1.70 | 2.12 | 2.47 | 2.64 | 2.93 | 3.18 | 3.48 | 3.73 | 4.02 | 4.29 |
| 13 | 1.73 | 2.14 | 2.51 | 2.67 | 2.96 | 3.22 | 3.52 | 3.77 | 4.07 | 4.35 |
| 14 | 1.75 | 2.16 | 2.54 | 2.70 | 3.00 | 3.25 | 3.56 | 3.82 | 4.12 | 4.40 |
| 15 | 1.78 | 2.19 | 2.58 | 2.73 | 3.03 | 3.29 | 3.59 | 3.87 | 4.16 | 4.45 |
| 16 | — | 2.21 | 2.61 | 2.76 | 3.07 | 3.33 | 3.64 | 3.92 | 4.21 | 4.50 |
| Nominal pressure | ||||||||||
| Rated burst pressure* | ||||||||||
Test data on file at Medtronic, Inc.
*Rated burst pressure: the maximum pressure to which a balloon is designed to be inflated (95% confidence that 99.9% will not fail below rated burst pressure). DO NOT EXCEED.
Nominal pressure: the pressure at which the balloon reaches its labeled diameter.