Triton Powered Surgical Instrument System
For Spinal, Cranial, and Orthopaedic Surgery

Indications, Safety, and Warnings

This website provides excerpts from our user manuals. It is important to consult the full system manuals for the most updated information. The most current version of each manual is located on our Medtronic Manual Library website.

Triton Electric High-Torque Handpiece

The Triton® Electric High-Torque Handpiece is capable of removing hard and soft tissue, drilling pilot holes, and driving screws, wires, and pins during spinal, cranial, and small-bone surgical procedures performed in an operating-room environment by surgeons trained in its use.

The electric Triton High-Torque Handpiece is part of the Integrated Power Console (IPC® System).

IPC System: Indications, Warnings, and Precautions

View indications, warnings, and precautions for the IPC System.

Triton Pneumatic Handpiece

Indications

The Triton Pneumatic Handpiece is intended for use during surgical procedures to cut hard tissue, bone, and soft tissue.

General Warning

Before using the Triton System, it is EXTREMELY important that you read the instruction manual.

Extreme caution must be observed when using Triton components near delicate neural and vascular structures. Improper use may cause injury to the patient or hospital staff.

We recommend in-service training by our representatives. If questions arise, please call one of our consultants at 800-643-2773 or 817-788-6400.

Precautions

Prior to each use, the entire Triton System must be inspected for proper operation. Check all components for gas leakage and return for servicing if necessary.

DO NOT OPERATE the system if the Triton hose is damaged.

Inspect the Triton hose before each use for any punctures, tears, or worn areas. Also check the metal bands at each end of the hose for looseness or wear. Do not attempt to repair a torn hose by wrapping tape around it; such taping could restrict the exhaust flow and cause the hose to rupture. If the Triton hose is damaged, send it in for repair.

Return the Triton hose at least once per year for routine inspection and service. Normal aging, repeated usage, and sterilization will gradually deteriorate the hose materials, causing cracks and breakage under pressure.

Attach the Triton handpieces to the hose and test for normal operation. NOTE: If the Triton hose does not insert easily into the handpieces, there are two possible problems:

  1.  The hose fitting or the handpiece fitting may be bent or damaged. Try another hose and handpiece to see if the damaged part can be identified. Return damaged parts for service.
  2. The o-ring in the handpiece connector may be dried out. It can be lubricated with a small amount of sterile petroleum jelly or sterile water (not saline).

Wear safety goggles or glasses when using the Triton System.

DO NOT initiate patient contact without testing instruments completely. Test for vibration, heat, etc., in the handpieces, attachments, and hoses.

ALWAYS change handpieces and attachments with instruments in the "SAFE" position.

DO NOT operate saw attachments or Compact Micro-Saws if blades cannot be firmly secured.

DO NOT use excessive force during surgery. Using excessive force may cause vibration or breakage, which may result in serious injury.

USE ONLY Triton saw blades with the Triton Handpiece saw attachments and Compact Micro-Saw Handpieces. Exact fit of blades is essential for stability and proper operation of equipment.

Warnings

DO NOT operate the Triton System unless you have been trained by an official Medtronic Midas Rex representative.

Prior to each use, the entire Triton System must be inspected for proper operation. Check all components for air leakage and return for servicing if leakage is noticed.

DO NOT initiate patient contact without testing instruments completely: test for vibration, heat, etc., in the handpiece, hose, and attachments.

Eye protection should be worn by operating room personnel and surgeons during use of the Triton System.

Following steam sterilization, the components must be allowed to cool to room temperature prior to use.

DO NOT operate the system if Triton hose is damaged.

DO NOT operate handpiece without making sure the proper attachment is used.

DO NOT use an attachment that gets hot or an attachment that is damaged in any way.

DO NOT operate the Sagittal Saw Attachment, Reciprocating Saw Attachment, Compact Micro-Sagittal Saw, Compact Micro-Oscillating Saw, or Compact Micro-Reciprocating Saw if blades cannot be firmly secured.

DO NOT use excessive force during surgery. Using excessive force may cause vibration or breakage, which may result in serious injury.

USE ONLY blades manufactured by Medtronic Midas Rex. Exact fit of blades is essential for stability and proper operation of equipment.

Always install a new blade prior to use.

 

Midas Rex Microsaws

Use Midas Rex® Microsaws to remove hard tissue and bone during surgical procedures. The Midas Rex Microsaws are part of the Integrated Power Console (IPC® System).

IPC System: Indications, Warnings, and Precautions

View indications, warnings, and precautions for the IPC System.

Midas Rex Legend EHS Stylus Surgical Drills

The Midas Rex® Legend EHS Stylus® Surgical Drill and Midas Rex Legend EHS Stylus Touch Surgical Drill are part of the Midas Rex Legend EHS-Electric High Speed- System (Legend EHS System) and the Integrated Power Console (IPC® System).

Indications

Legend EHS System:

The Midas Rex Legend EHS–Electric High Speed System is intended for use in surgical procedures for the following medical applications:

  • Neurosurgical
  • Spine
  • Orthopaedic surgery
  • General surgery
  • ENT
  • Maxillofacial
  • Craniofacial

Contraindications

Legend EHS System:
None.

Warnings

Legend EHS System:

DO NOT interchange components of the Legend EHS System with components from the Midas Rex Classic Motor System, the Midas Rex III Motor System or the Mednext® Bone Dissecting System and their associated attachments and dissecting tools.

DO NOT operate the Legend EHS System in a Magnetic Resonance Imaging Suite.

DO NOT use Legend EHS Motors or Legend Attachments before following proper cleaning and sterilization protocols.

DO NOT dissect bone without adequate irrigation to prevent thermal necrosis.

DO NOT operate the Legend EHS Motors without protective eyewear.

DO NOT attempt to run the Legend EHS motors immediately after autoclaving. Allow an adequate "cool down" period (Typically 1 hour).

DO NOT use the Legend EHS System in the presence of flammable anesthetics.

DO NOT use Legend EHS System components if damage is apparent or if components do not run properly. The Legend EHS System must be inspected for damage prior to each use.

  • Conduct a visual inspection of the cables for cracks, tears or corrosion.
  • Check attachments for proper appearance. Install attachment and dissecting tool, then briefly run motor.
  • Check motor for overheating.
  • Check attachment for overheating.
  • Check dissecting tool for flail.
  • Check for bent or missing pins in cable connectors

DO NOT attempt to use an overheating motor or attachment.

Consult the Legend Bone Mill product insert before use with the Legend EHS System.

For metal transection, observe the following safety precautions:

  1. Protective eyewear is essential.
  2. Irrigate copiously to cool the cutting surfaces.
  3. Protect the wound site from metal debris.
  4. Use a clamp or grasping device to control loose fragments during transection of any metal component.

Legend EHS Motors or Legend Attachments which fail due to extended use may as a result of such failing allow a component to detach and fall from the motor or attachment and may cause patient injury.

Electrical contacts must be properly dried prior to use.

DO NOT attempt to change a dissecting tool or attachment while the motor is running, or when the motor or attachment is in an overheated state.

DO NOT use a Legend EHS System if the motor continues to run after releasing the foot pedal. If a Legend EHS System requires servicing or refurbishing, return system to Medtronic Powered Surgical Solutions.

DO NOT immerse the Legend EHS System Motor or Cables.

DO NOT place motor, attachment and dissecting tool on the patient or in an unsecured location during surgery.

DO NOT open or disassemble the console. There are no user serviceable parts inside. Refer all servicing to Medtronic Powered Surgical Solutions.

DO NOT immerse the console.

DO NOT sterilize.

DO NOT connect any cables or equipment to the console unless the cables and equipment are approved for use by Medtronic Powered Surgical Solutions as documented in the appropriate Instructions for Use and Classified by UL for use with this system.

DO NOT use a dissecting tool without the appropriate attachment as injury may occur to the patient and/or operator.

DO NOT use a Legend Attachment if any part of the attachment appears to be bent, loose, missing or damaged. If a Legend Attachment requires servicing or refurbishing, return attachment to Medtronic Powered Surgical Solutions.

DO NOT force or modify any attachment or dissecting tool to fit a Legend EHS or Stylus Motor.

Consult the Legend Bone Mill product insert before use with the Legend EHS System.

Heavy side loads and/or long operating periods may cause overheating of the attachment to the point where the attachment is uncomfortable to hold.

  1. Never place an overheated attachment on the patient or patient draping during surgery.
  2. Mitigate the overheating by discontinuing use and rest the attachment by using intermittently, or wrap the attachment interface with a moist sterile towel.
  3. If the attachment is passed off, the receiver of the attachment must handle the attachment cautiously.

Remove Legend Footed Attachments cautiously and slowly as per instructions to avoid injury to the operator.

DO NOT use a dissecting tool without the appropriate attachment as injury may occur to patient, operator and/or operating room staff.

DO NOT use excessive pressure, such as bending or prying, on attachments or dissecting tools. This may cause tool to bend or break and cause injury to patient, operator and/or operating room staff.

DO NOT use an attachment and dissecting tool combination if flail or vibration occurs.

DO NOT use metal cutting dissecting tools on bone.

DO NOT use a twist drill or contra-angle dissecting tool at an operating speed over 62,000 rpm.

DO NOT use a dissecting tool if the sterile packaging is damaged.

If a dissecting tool package is opened, but the tool is not used or contaminated, the tool can be re-sterilized. Remove tool from original packaging and place into an approved autoclave package. Steam sterilize as follows:

  • High Vacuum Steam 270°F (132°C) for 5 minutes
  • Gravity Displacement 270°F (132°C) for 15 minutes

The re-sterilized tool must be used promptly following re-sterilization. If rust or corrosion is encountered after re-sterilization, do not use the re-sterilized tool.

Dissecting tool flutes are sharp and can perforate surgical gloves. Always handle with care. Do not spin, twist, or pull a tool by grasping the flutes. Gently pull on dissecting tool shaft to ensure that it is fully seated and properly installed in motor.

DO NOT use the Legend EHS foot control to operate any systems other than Legend EHS.

DO NOT open or disassemble the Foot Control. There are no user serviceable parts inside. Refer all servicing to Medtronic Powered Surgical Solutions.

DO NOT sterilize the Foot Control.

DO NOT use cables with cracks, tears or corrosion.

DO NOT use cables with bent or missing connector pins.

DO NOT kink EHS cables.

DO NOT use power cable with cracks, tears or corrosion.

DO NOT use power cable with bent or missing connector pins.

DO NOT use any other components except for Midas Rex Legend® Attachments, Legend Dissecting Tools, and Legend accessories in combination with Legend EHS Motors. Use of devices other than Legend EHS System or associated parts or components may cause injury to the patient, operator and/or operating room staff. Use of devices other than Legend EHS System or associated parts or components will damage the motor and attachments and will void manufacturer’s limited warranty.

DO NOT use any other components except for Midas Rex Legend® Attachments, Legend Dissecting Tools and Legend accessories in combination with a Legend EHS or Stylus Motor. Use of devices other than Legend EHS Systems or associated parts or components may cause injury to the patient, operator and/or operating room staff. Use of devices other than Legend EHS Systems or associated parts or components will damage the motor and attachments and will void manufacturer’s limited warranty.

DO NOT re-use or re-process a dissecting tool that has already been used. Using a dissecting tool that has already been used may cause injury to the patient, operator and/or operating room staff. Using a dissecting tool that has already been used may affect motor performance and will void manufacturer’s limited warranty.

DO NOT use an attachment and dissecting tool combination if flail or vibration occurs.

DO NOT use a twist drill or contra-angle dissecting tool at an operating speed over 62,000 rpm.

Do not soak/submerge Legend devices.

Do not use ultrasound to clean Legend devices.

Do not use chlorine based or corrosive cleaning agents such as bleach, lye, acetone, sodium hypochlorite/bleach, sodium hydroxide, formic acid, or solutions containing glutaraldehyde.

The use of a washer-disinfector for cleaning may cause a premature degradation in performance.

Allow an adequate cooling period after steam sterilization.

Do not steam or EO sterilize the Legend attachment cleaning nozzle.

Do not use Pana Spray on variable exposure attachments, as it may cause the attachment to overheat.

Surgeons should familiarize themselves with the performance of dissecting tools before use, and should explore the effect of various levels of tool exposure on dissection stability. If the tool exhibits excessive chatter, vibration, or movement, decrease the tool exposure.

Dissecting tool size and geometry may contribute to excessive vibration at certain speeds. Increase or decrease speed by adjusting the foot/finger control, or by changing the operating pressure or console speed setting. If necessary, use a different dissecting tool.

Do not use the Variable Exposure Attachment if the TUBE adjustment ring spins freely or fails to click into place with each adjustment, as the exposure may change without warning.

Do not use the end of the tube as a depth gauge or depth stop.

DO NOT use a Legend Attachment if any part of the attachment appears to be bent, loose, missing or damaged. Excessive pressure or improper handling, such as bending or prying, with the attachment or dissecting tool may cause injury to patient, operator and/or operating room staff. If a Legend Attachment requires servicing or refurbishing, return attachment to Medtronic Powered Surgical Solutions.

Remove Legend Footed Attachments cautiously and slowly to avoid injury to the operator.

DO NOT use metal cutting dissecting tools on bone.

For metal transection, observe the following Safety Precautions:

  • Eyewear protection is essential.
  • Irrigate copiously to cool the cutting surfaces.
  • Protect the wound site from metal debris.
  • Use clamp or grasping device to control loose fragments during transection of any metal component.

Dissecting tool flutes are sharp and may perforate surgical gloves. Dissecting tools for the Telescoping Attachment may be grasped with a hemostat to aid in installation and removal.

CONSULT the cranial perforator device labeling for the recommended speed specifications.

Indications

IPC System:

The IPC® System is indicated for the incision / cutting, removal, drilling, and sawing of soft and hard tissue and bone, in Head & Neck / ENT (Otologic, Neurologic, Neurotologic, Sinus, Rhinologic, Nasopharyngeal / Laryngeal), Oral / Maxillofacial, and Plastic / Reconstructive / Aesthetic, surgical procedures.

Contraindications

IPC System:
None.

Warnings

WARNINGS

System Warnings
W1It is important that the IPC operator be familiar with the system User’s Guide, its precautions, procedures and safety issues.
W2Do not use the IPC System in the presence of flammable anesthetics. Avoid potential ignition or explosion of gases
W3When not operating handpiece, eliminate accidental foot control activation. Control energy to and through the handpiece to prevent unintended tissue, bone, or nerve resection.
W4Disconnect power to the IPC before cleaning the unit to avoid electrical macro shock.
W5Do not attach unapproved components to the IPC to avoid electrical macro shock.
W6To avoid the risk of electrical shock, achieve electrical grounding reliability with proper connections. Connect the IPC to hospital grade receptacles only.
W7This medical device complies with EN60601-1-2 safety standard for electromagnetic compatibility, requirements and test. However, if this equipment is operated in the presence of high levels of electromagnetic interference (EMI) or highly sensitive equipment, interference may be encountered and the user should take whatever steps are necessary to eliminate or reduce the source of the interference. Diminished performance may lengthen operating time for anesthetized patient.
W8Medical Electrical Equipment needs special Precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this Guide.
W9Portable and mobile RF communications equipment can affect Medical Electrical Equipment.
W10Do not operate the IPC System in the presence of Magnetic Resonance Imaging devices.
W11Use of accessories and cables other than those specified and sold by Medtronic may result in increased emissions and decreased immunity of this unit.
W12The IPC should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the IPC should be observed to verify normal operation in the configuration in which it will be used.
W13Do not attempt to run the IPC System handpiece immediately after autoclaving. Allow an adequate “cool down” period (Typically 1 hour).
W14Consult the Legend® Bone Mill product insert before use with the Integrated Power Console.
W15For metal transection, observe the following safety precautions:
 
W15aEye wear protection is essential.
W15bIrrigate well to cool the cutting surfaces.
W15cProtect the wound site from metal debris.
W15dUse a clamp or grasping device to control loose fragments during transection of any metal component.
W16Do not operate the IPC System without eye protection.
W17All service must be performed by Medtronic qualified personnel only
W18Repair and/or modification to the IPC by anyone other than qualified service personnel may significantly compromise the unit’s ability to perform effectively and/or void the equipment warranty.
Component Warnings
W19Do not use any parts other than Medtronic system components as damage or substandard performance could result.
W20Always inspect the components before and after use for any damage. If damage is observed, do not use damaged part until it is repaired or replaced. Damaged parts may deposit metal shavings on surgical site.
W21When precise location of blade tip is required, engage the rotation lock on the handpiece, then calibrate and verify the blade tip on Image Guided Surgery (IGS) system. Always lock M4 handpiece when driving non-rotatable blades to maintain their IGS calibration.
W22Employ visualization, including use of imaging techniques (e.g., fluoroscopy, image guided surgery) when using rotating powered accessories. Discontinue powered application in the event of lack of visualization of surgical site..
W23Midas Rex® Variable Exposure attachments
Surgeons should familiarize themselves with the performance of dissecting tools before use, and should explore the effect of various levels of tool exposure on dissection stability. If the tool exhibits excessive chatter, vibration, or movement, decrease the tool exposure.
W24Motors and attachments may fail due to extended use and allow a component to detach and fall from the motor or attachment, causing patient injury.
W25Electrical contacts must be dry prior to use.
W26Heavy side loads and/or long operating periods may cause the device to overheat.
W27Do not use an overheated device, as it may cause thermal injury to the patient or operator.
W28Use adequate irrigation. The use of a tool without irrigation may cause an inordinate amount of heat buildup resulting in a thermal injury to tissue. Depending on the amount of irrigation used, the drill bits can achieve temperatures in excess of 50° C.
W29Do not attempt to change a dissecting tool or attachment while the motor is running, or when the motor or attachment is in an overheated state.
W30Do not immerse the system components.
W31Do not place motor, attachment and tool on the patient or in an unsecured location during surgery.
W32A system that is not functioning properly should not be used until all necessary repairs have been made and the unit is tested to ensure that it is functioning in accordance with Medtronic specifications.
W33Match the nomenclature and color code on the tool packaging to the same nomenclature and color code on the Attachment.
W34Make sure that the attachment is still in the locked position after each adjustment of the tool exposure, as attempting to increase the tool exposure too far, may result in the attachment accidentally being unlocked.
W35Midas Rex® Legend EHS® Motor and Midas Rex Legend EHS Stylus® Motor should only be operated when the attachment is in the locked position.
W36Smoke may be generated if attachment is not in the locked position.
W37The Legend EHS motors will not run properly unless the attachment is in the locked position.
W38DO NOT change accessory with handpiece running to prevent laceration of user and cross-contamination through compromised glove.
W39Remove Legend Footed Attachments cautiously and slowly as per instructions to avoid injury to the operator.
W40DO NOT modify accessories used with the handpiece. Performance could be diminished with modified accessories.
W41The safe use of the Endo-Scrub® 2 System in procedures where surgical lasers are also employed has not been clinically demonstrated.
W42In order to ensure compliance with requirements of IEC 60601-1, use a Medtronic approved power cable
W43To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
W44Keep NIM® Muting Probe cable away from IPC system cables.
Disposable Warnings
W45Verify reusable device was sterilized prior to use. If not sterilized, do not use.
W46Tools are available for resection of soft tissue and bone for surgical procedures. Use of tools depends on the intended application and patient needs. Sharp-cutting powered tools induce bleeding and removal of significant tissue and bone.
W47Use methods at the operative site to control bleeding that do not compromise patient safety during at-risk surgery.
W48Always keep the cutting tip of the tool away from fingers and loose clothing. Prevent laceration of user and cross-contamination through compromised glove.
W49Operate the tool only after the appropriate anatomical landmarks and the intended surgical site have been confirmed.
W50Use care in application of the moving cutting end to only appropriate anatomical landmarks and the intended surgical site when using powered accessories.
W51Insertion of metal objects in accessory tip may cause the accessory to break leaving fragments in the wound. The fragments may be difficult to remove, causing irritation, inflammation and foreign-body response at surgical site.
W52Bending or prying may break the accessory, causing harm to patient or staff.
W53Do not use excessive force to pry or push bone with the attachment or tool during dissection.
W54A tool’s size and geometry may create excessive vibration at certain speeds. Increase or decrease speed on console. Change to a new tool to prevent unintended tissue removal from patient.
W55Test for wobble at desired speed prior to use. Discontinue use of accessory if tip begins to wobble and replace accessory to prevent unintended tissue removal from patient.
W56Eccentricity of the tool can cause tool vibration and may result in excess tissue and bone destruction and hearing damage.
W57Excessive noise from the tool when drilling close to the cochlea or ossicular chain may cause hearing damage.
W58CONSULT the cranial perforator device labeling for the recommended speed specifications.
W59Tools with L” identification are longer tools intended for light bone dissection. The increased tool head/stem configuration may affect dissection stability.
W60Tool flutes are sharp and may perforate surgical gloves. Tools may be grasped with a hemostat to aid in installation and removal.
W61DO NOT attempt to resharpen used tools. Worn tools should be replaced with new ones frequently to ensure effective cutting and control.
W62Carefully inspect tool both prior to and following each use for signs of excessive wear, fragmentation, eccentricities or other defects. Replace any suspicious tools with a new one prior to use.
W63Excessive pressure applied to bur may cause bur fracture. Should a tool fracture in use, extreme care must be exercised to ensure that all fragments of the tool are retrieved and removed from the patient. Unremoved tool fragments may cause tissue damage to the patient.
W64Do not use metal-cutting tools on bone.
W65Use only rotary tools specifically designed for use with this drill system.
W66When using non-rotatable tools, ensure rotation lock is engaged to prevent inadvertent rotation.
W67The use of powered reciprocating instruments may result in vibration-related injury.
W68Powered blades should be operated in the oscillate mode only. Operating in the forward mode may cause damage to the blade.
W69Do not attempt to sterilize disposable devices. The disposables are packed sterile and are not intended for repeat use. To prevent contamination, use only once.
W70Any tubing or other tip protectors used during shipping must be removed prior to cleaning and sterilization.
W71Do not use accessory if package is opened or damaged. Broken seal offers no protection against cross-contamination.
W72Properly dispose of single-use devices removed from sterile packages. Devices lose sterility upon removal from packaging.
W73Do not use dull, damaged or bent tools. Use of dull tools can reduce handpiece effectiveness and cause the handpiece temperature to increase.
W74T&A Blades: Gently remove the inner tube from the outer tube. The inner tube may elongate upon removal from the outer tube. If this occurs, the inner tube may not lock properly into the handpiece or the blade may not work properly.
W75T&A Blades: Rotate the inner tube when removing and inserting it in the outer tubes to prevent damage to the internal seal. If the seal is damaged, the blade will leak at the handpiece.
W76Always ensure that the drill is securely engaged into the handpiece prior to operating the system.
W77Always examine operation of each tool in a handpiece before use.
W78Powered burs and drills should be operated in the forward mode only.
W79This system requires insulated connectors for the Straightshot® M4 Microdebrider, Straightshot Magnum® II Microdebrider, Straightshot III Microdebrider, Midas Rex® SC1, Visao®, or Skeeter® handpieces and the Multi Function Foot Control Unit.
W80Sterilize and dry reusable device before storing the system. Decrease likelihood of cross-contamination with timely sterilization.
W81After each procedure, properly clean all reusable system components.
W82Auxiliary Power Outlet with protective cover is for use with the Hydrodebrider™, or Bone Mill consoles only.

 

Last updated: 26 Feb 2013