Verifying a SureScan® Pacing System

Steps to easily verify the presence of a SureScan Pacing System

In order to utilize the MRI capabilities of the SureScan Pacing System, first confirm that the patient's device and leads are
MR conditional.

1. The patients device ID card identifies the patient as having a Revo MRI or Advisa MRI SureScan Pacing System.

Revo MRI Indentification Card


Advisa MRI Indentification Card


2. Call Medtronic at 1 (877) MRI-7677 to verify the patient's pacing system.

3. Use the patients records to verify a complete SureScan pacing System has been implanted.

• Confirm that the patient does not have other implanted medical devices, active, passive or abandoned, that are located within 10 cm of the implanted pacing system components. The interaction between pacing system components and other implantable devices within 10 cm during MRI scans has not been evaluated by Medtronic.
• Scanning patients who have multiple MR Conditional devices present is acceptable as long as the MR labeling conditions for all implants can be satisfied.

4. Easily identifiable radiopaque SureScan icon on the device and lead confirms the presence of a complete system.

System Status

Revo MRI Device Verification

Device Verification

Unique radiopaque symbol on the device
above the three-letter code

CapSureFix MRI 5086MRI lead

Lead Verification

Unique radiopaque helix on the leads

Advisa MRI Device Verification

Device Verification

Unique radiopaque symbol on the device above the three-letter code

5. Medtronic programmer verifies the device. On-screen programming clearly outlines necessary steps to ensure safe scanning.

Tutorials for cardiology and radiology

The above tutorials provide practical information on identifying the Revo MRI SureScan Pacing System and MR-conditional criteria.