Getting a Replacement – Drug Pumps
As someone who has chosen a Medtronic drug pump for chronic pain, you know first-hand the difference it can make. We want to ensure that you continue to benefit from your pain therapy.
Battery Life: SynchroMed II
The Medtronic SynchroMed® II pump uses a battery to keep the drug flowing through the pump and catheter. The typical battery life of your SynchroMed II pump will range between 5 and 7 years, depending on how much medicine you’re programmed to receive each day. Your doctor will closely monitor the performance and battery levels of your programmable pump at each of your refill appointments.
It is important to know that the SynchroMed II pump will last a maximum of 7 years. The automatic shut-off feature helps ensure that the pump works correctly to provide pain therapy.
If you have a SynchroMed II pump, you will need to have your pump surgically replaced with a new pump before it automatically shuts off. This requires surgery.
Battery Life: SynchroMed EL
Like the SynchroMed II, the SynchroMed EL pump (most pumps implanted before July 2004) will sound an alarm when the battery is low. However, it will not automatically shut down. Depending on the rate at which you receive medication, the low battery alarm on most SynchroMed EL pumps will sound after 5 to 8 years of use.
If you have a SynchroMed EL pump, you will need to have your pump surgically replaced with a new pump before it automatically shuts off. This requires surgery.
When to Schedule a Replacement
The SynchroMed II pump (the pump most often implanted after July 2004) has an elective replacement indicator (ERI) alarm. This alarm sounds when your pump is nearing End of Service. If the pump is not replaced after the ERI alarm sounds, the pump will continue to operate for a while, but will stop after 90 days. A stopped pump results in a loss of therapy that can lead to an underdose or withdrawal symptoms.
It is best not to wait for this alarm to sound before scheduling your pump replacement. Medtronic recommends that you schedule your replacement about 6 months prior to the automatic shut off at 7 years so there is no interruption in your therapy.
If you have a SynchroMed EL drug pump, contact your doctor as soon as you hear the low-battery alarm to schedule a replacement.
Be sure to tell your doctor right away if you experience any unusual symptoms, side effects, or changes in your condition.
For more information about your pump’s service life, contact your doctor.