|
MRI Guidelines for Medtronic Deep Brain Stimulation Systems
Download the MRI Guidelines for Medtronic Deep Brain Stimulation Systems information in PDF format.
Medtronic®, Activa®, Kinetra®, Soletra®, and Itrel® are registered trademarks of Medtronic Inc.
DBS is a trademark of Medtronic, Inc.
Introduction
It is important to read this manual in its entirety before conducting an
MRI examination on a patient with any implanted Activa System
component. Contact Medtronic if you have any questions.
Due to the number and variability of parameters that affect MRI
compatibility, the safety of patients or continued functioning of Activa
Systems exposed to MRI cannot be absolutely ensured. MRI
systems generate powerful electromagnetic fields that can produce
a number of interactions with implanted components of the Activa
neurostimulation system. Some of these interactions, especially
heating, are potentially hazardous and can lead to serious injury or
death. However, with appropriate control measures, particularly with
respect to the selection of MRI parameters and RF coils, it is
generally possible to safely perform an MRI head scan on an Activa
patient. In addition, Activa System components can affect the MRI
image, potentially impacting the diagnostic use of this modality. The
following information describes the potential interactions and control
measures that should be taken to minimize the risks from these
interactions.
Contraindication: Implantation of an Activa Brain Stimulation
System is contraindicated for patients who will be exposed to
Magnetic Resonance Imaging (MRI) using a full body transmit
radio-frequency (RF) coil, a receive-only head coil, or a head
transmit coil that extends over the chest area. Performing MRI
with this equipment can cause tissue lesions from component
heating, especially at the lead electrodes, resulting in serious
and permanent injury including coma, paralysis, or death.
Warnings:
- Do not conduct an MRI examination on a patient with any
implanted Activa System component until you read and
fully understand all the information in this manual. Failure
to follow all warnings and guidelines related to MRI can
result in serious and permanent injury including coma,
paralysis, or death.
- In-vitro testing has shown that exposure of the Activa
neurostimulator system to MRI at parameters other than
those described in this guideline can induce significant
heating at the lead electrodes or at breaks in the lead.
Excessive heating may occur even if the lead and/or
extension are the only part of the Activa System that is
implanted. Excessive heating can result in serious and
permanent injury including coma, paralysis, or death.
- MRI examinations of patients with an implanted Activa
System should only be done if absolutely needed and then
only if these guidelines are followed. MRI should not be
considered for Activa patients if other potentially safer
diagnostic methods such as CT, X-ray, ultrasound, or other
methods will provide adequate diagnostic information.
- A responsible individual with expert knowledge about MRI,
such as an MRI radiologist or MRI physicist, must assure
all procedures in this guideline are followed and that the
MRI scan parameters, especially RF specific absorption
rate (SAR) and gradient dB/dt parameters, comply with the
recommended settings, both for the pre-scan (tuning) and
during the actual MRI examination. The responsible
individual must verify that parameters entered into the MRI
system meet the guidelines in this manual.
- Do not conduct an MRI examination if the patient has any
other implants or limiting factors that would prohibit or
contraindicate an MRI examination.
Cautions:
- The neurostimulator, especially those without filtered
feedthroughs such as the Itrel II Model 7424, may be reset
or potentially damaged when subjected to an MRI
examination. If reset, the neurostimulator must be
reprogrammed. If damaged, the neurostimulator must be
replaced.
- MRI images may be severely distorted or image target
areas can be completely blocked from view near the
implanted Activa System components, especially near the
neurostimulator. If the MRI targeted image area is near the
neurostimulator, it may be necessary to move the
neurostimulator to obtain an image, or use alternate
imaging techniques. Do not remove the neurostimulator
and leave the lead system implanted as this can result in
higher than expected lead heating.
- Carefully weigh any decision to perform magnetic
resonance imaging (MRI) examinations on patients who
require the neurostimulator to control tremor. Image
quality during MRI examinations may be reduced,
because the tremor may return when the neurostimulator
is turned off.
- If possible, do not sedate the patient so that the patient
can provide feedback of any problems during the
examination.
- Monitor the patient during the MRI examination. Verify that
the patient is feeling normal and is responsive between
each individual scan sequence of the MRI examination.
Discontinue the MRI immediately if the patient becomes
unresponsive to questions or experiences any heating,
pain, shocking sensations/uncomfortable stimulation, or
unusual sensations.
Note: The MRI guidelines provided here may significantly extend the
MRI examination time or prevent some types of MRI examinations
from being conducted on Activa patients.
General Information on MRI
An MRI system produces three types of electromagnetic fields that
may interact with implanted neurostimulation systems. All three of
these fields are necessary to produce an MRI image. Each of these
fields can also produce specific but different types of interactions with
implanted neurostimulator systems. These fields include:
- Static Magnetic Field. This is a steady state non-varying
magnetic field that is normally always ON, even when no scan
is underway. In a 1.5 Tesla MRI system, the static magnetic
field is approximately 30,000 times greater than the magnetic
field of the earth.
- Gradient Magnetic Field. This is a low-frequency pulsed
magnetic field that is only present during a scan. The gradient
magnetic field can induce voltages onto the lead system that
may result in unintended stimulation or functional interactions
with the neurostimulator.
- RF Field. This is a pulsed radio frequency (RF) field that is only
present during a scan. It can be produced by a variety of
transmission RF coils such as a whole body transmit coil or an
extremity coil such as a transmit/receive head coil. Only a
transmit/receive head coil should be used as the other RF coils
can expose more of the lead system to RF energy, thereby
increasing the risk of excessive heating and thermal lesions
possibly resulting in coma, paralysis, or death.
MRI Interactions with Implanted Activa Systems
MRI/neurostimulation system interactions are various, and the risk to
the patient can range from minimal to severe. These interactions
include the following:
Heating - The MRI RF field induces voltages onto the lead system
that can produce significant heating effects at the lead-electrode tissue
interface or at the location of any breaks in the neurostimulator
lead system. Component heating from the MRI RF field is the most
serious risk from MRI exposure. Failure to follow these MRI
recommendations can result in thermal lesions possibly resulting in
coma, paralysis, or death.
Magnetic Field Interactions - Magnetic field interactions such as
force and torque effects are produced by the static magnetic field.
Any magnetic material will be attracted to the static magnetic field of
the MRI. The force and torque effects may produce movement of the
neurostimulator that can be uncomfortable to the patient, open a
recent incision, or both. Activa System components are designed
with minimal magnetic materials.
Induced Stimulation - Gradient magnetic fields may induce
voltages onto the lead system that may cause unintended
stimulation. The voltage of the induced stimulation pulses is
proportional to the time rate of change (dB/dt) of the gradient pulses,
the effective loop area created by the neurostimulator lead system,
and the location of the lead system with respect to the gradient coils
of the MRI.
Effects on Neurostimulator Function - The static, gradient, and
RF fields of the MRI may affect the neurostimulator operation and
programming. The static magnetic field may cause the
neurostimulator to turn ON or OFF if the neurostimulator uses a
magnetically controlled switch that allows the patient to control
stimulation by the application of a handheld magnet. Additionally, the
MRI RF, static, and gradient fields may temporarily affect or disable
other functions, such as telemetry or stimulation pulses. Parameters
will need to be reprogrammed if the MRI causes a POR (Power On
Reset) of the neurostimulator.
Image Artifacts and Distortion - The neurostimulation system
components, particularly the neurostimulator, can cause significant
imaging artifacts and/or distortion of the MRI image, particularly if the
neurostimulator components contain magnetic material. The
neurostimulator can cause the MRI image to be completely blocked
from view (i.e., signal loss or signal "void") or severely distorted within
several inches of the neurostimulator.
MRI Procedure
Scope
These MRI/neurostimulator exposure guidelines apply to Activa
Systems comprising combinations of the following components:
-
Neurostimulator Models: Itrel II 7424, Soletra 7426,
Kinetra 7428
- Lead Extension Models: 7495, 7482
- Lead Models: DBS 3387, 3389
Supervision
A responsible individual such as an MRI radiologist or MRI physicist
must assure these procedures are followed. If the MRI is operated by
an MRI technician, it is strongly recommended the responsible
individual verifies that the MRI recommendations are followed.
Preparation
Do the following prior to performing an MRI examination on an Activa
patient:
- Inform the patient of the risks of undergoing an MRI.
- Check if the patient has any other implants or conditions that
would prohibit or contraindicate an MRI examination. Do not
conduct an MRI examination if any are found.
- Verify that all proposed MRI examination parameters comply
with the "MRI Operation Settings." If not, the
parameters must be modified to meet these requirements. If
this cannot be done, do not perform an MRI.
- If the patient has implanted leads but does not have an
implanted neurostimulator, perform the following steps:
- Wrap the external portion
of the leads/percutaneous extensions with insulating material.
- Keep the
external portion of the leads/percutaneous extensions out of contact
with the patient.
- Keep the external leads/percutaneous extensions straight,
with no loops, and running down the center of the head
coil.
- If the patient has an implanted neurostimulator, perform the
following steps:
- Review the neurostimulator with a clinician programmer
and print out a copy of the programmed parameters for
reference.
- Test for possible open circuits by measuring impedance
and battery current on all electrodes in unipolar mode (see
Table 1). If an open circuit is suspected, obtain an x-ray to
identify whether the open circuit is caused by a broken lead
wire. If a broken lead wire is found, do not perform an MRI.
| Table 1. Measurement Values Indicating Possible Open Circuits |
| Neurostimulator |
Impedance |
Battery Current |
| Itrel II Model 7424 |
>2000 Ω |
<10 μA |
| Soletra Model 7426 |
>2000 Ω |
<10 μA |
| Kinetra Model 7428 |
>4000 Ω |
<15 μA |
-
- Warning: An MRI procedure should not be performed in a
patient with an Activa System that has a broken lead wire
because higher than normal heating may occur at the break
or the lead electrodes which can cause thermal lesions.
These lesions may result in coma, paralysis, or death.
- If the Activa System is functioning properly and no broken
lead wires are found, program the neurostimulator to the
settings provided in Table 2.
| Table 2. Recommended Neurostimulator Settings for MRI |
| Parameter |
Setting |
| Stimulation output |
OFF (all programs) |
| Stimulation mode |
Bipolar (all programs) |
| Amplitude |
0 Volts (all programs) |
| Magnetic (reed) switch |
Disabled (Kinetra Model 7428 only) |
| Other parameters |
Do not change |
MRI Operation Settings
Prior to the MRI examination, a responsible individual such as an
MRI radiologist or MRI physicist must assure the examination will be
conducted according to the following MRI requirements. If standard
MRI pulse sequences will be used, they must meet these
requirements. If they do not, the pulse parameters must be adjusted
so that they comply with these requirements:
Warning:
In-vitro testing has shown that exposure of the
Activa System to MRI under conditions other than described
in this guideline can induce excessive heating at the lead
electrodes or at breaks in the lead to cause lesions. These
lesions may result in coma, paralysis, or death. -
Use only a 1.5 Tesla horizontal bore MRI (do not use open
sided or other field strength MRI systems).
- Use only a transmit/receive head coil.
Contraindication: Implantation of an Activa Brain
Stimulation System is contraindicated for patients who
will be exposed to Magnetic Resonance Imaging (MRI)
using a full body transmit radio-frequency (RF) coil, a
receive-only head coil, or a head transmit coil that
extends over the chest area. Performing MRI with this
equipment can cause tissue lesions from component
heating, especially at the lead electrodes, resulting in
serious and permanent injury including coma, paralysis,
or death.
- Enter the correct patient weight into the MRI console to assure
the head SAR is estimated correctly.
- Use MRI examination parameters that limit the displayed average
head SAR to 1/10 (0.1) W/kg or less for all RF pulse sequences unless the applied SAR is known. If known, an applied SAR up to 1/10 (0.1) W/kg may be used.
Warnings:
- Ensure the SAR value is the value for head SAR.
Some MRI systems may only display SAR, whole
body SAR, or local body SAR. Make sure the value
being limited is for head SAR. Excessive heating may
occur if the wrong SAR value is used.
- If MRI parameters must be manually
adjusted after the initial automatic MRI prescan, do not make any
adjustments that will increase the SAR value. Some
MRI machines may not automatically update the
displayed SAR value if manual adjustments are
made. This may lead to higher than expected
temperature increases in the Activa System,
particularly at the lead electrodes.
-
Limit the gradient dB/dt field to 20 Tesla/second or less.
Note: The recommendations provided are based on in-vitro testing
and should result in a safe MRI examination of a patient with an
implanted Medtronic Activa System. However, due to the many
variables that affect safety, Medtronic cannot absolutely ensure
safety or that the neurostimulator will not be damaged. The user of
this information assumes full responsibility for the consequences of
conducting an MRI examination on a patient with an implanted Activa
System.
Prior to the MRI Examination
Prior to the scan examination, the responsible
individual must verify the MRI examination parameters comply with these guidelines.
- Patients with implanted Activa Systems should be informed of
the risks of undergoing an MRI.
- If possible, do not use sedation so the patient can inform the
MRI operator of any heating, discomfort, or other problems.
- Instruct the patient to immediately inform the MRI operator if
any discomfort, stimulation, shocking, or heating occurs during
the examination.
During the MRI Examination
-
Monitor the patient both visually and audibly. Check the patient
between each imaging sequence. Discontinue the MRI
examination immediately if the patient is unable to respond to questions or reports any problems.
- Conduct the examination using only the MRI pulse sequence
that the MRI radiologist or physicist has confirmed meets the
MRI requirements above.
Post MRI Examination Review
-
Verify that the patient is feeling normal.
- Verify that the neurostimulator is functional.
- Reprogram the neurostimulator to pre-MRI settings.
|
|