1-2-3 Follow-Up with the Kappa® 900 Pacing System
Depending on your patients condition, follow-up with the Kappa 900 Pacing System can be a few simple steps:
 |
1. Tap Auto-Identify to automatically:
- ID and interrogate the device
- Print the Initial Interrogation report, including the Arrhythmia Summary
|
| |
2. Review the following:
- Longevity and Battery Voltage
- Ventricular Capture Management results
- Sensing Assurance results
- Chronic Lead Impedance
- Arrhythmia Summary Report
|
 |
3. Tap the Fast Path button next to the Checklist icon, then:
- Perform the Atrial Strength Duration Test
- Reprogram outputs (if called for by test)
|
 |
4. Tap the Fast Path button next to the Checklist icon, then:
- Select and print applicable reports (usually only the Strength Duration Threshold Test Report and/or Final Report)
|
 |
5. If the patient is symptomatic and the data indicates the need for other tests, go to the Tests icon and select the appropriate test.
|
| |
6. If your clinic calls for a Magnet/Non-Magnet ECG strip as part of its protocol, create a new Follow-Up checklist by adding this test to the Quick Look Follow-Up. |
|