InSync II Marquis™ CRT-D Device
The Medtronic InSync II Marquis™ cardiac resynchronization and ICD device offers patient-tailored CRT options and effective ICD protection in conjunction with powerful heart failure management and patient safety features.
Key Features
Fast, Effective ICD Therapy Combined with Patient-Tailored Cardiac Resynchronization Therapy
- 30 Joules high energy output in 5.9 to 7.5 seconds throughout the life of the device.
- Painless antitachycardia pacing (ATP) therapy to terminate 3 out of 4 episodes of fast ventricular tachycardia (VT).1
- Enhanced PR Logic dual-chamber detection to maintain high sensitivity for ventricular tachyarrhythmias while improving discrimination for supraventricular tachycardias. Offers 95.2% accuracy (Positive Predictive Value) in delivered therapies.2
- Separately programmable ventricular outputs for adapting pacing outputs to unique patient needs.
- Special pacing functions to manage heart failure patients who also experience atrial fibrillation.
- Programmable with the Medtronic CareLink® Programmer
Powerful Heart Failure Management and Advanced Patient Alert Capabilities
- Heart Failure Management Report showing 14-month trends of key heart failure patient monitoring categories such as AT/AF Burden, ventricular rate during AF, heart rate variability, patient activity and night/day heart rates. Observation highlight heart failure related parameters outside of typical ranges.
- Cardiac Compass® Report with additional information about ventricular arrhythmia detection and therapies.
- Rate Histogram Report provides information about % pacing during episodes of AF, sinus rate profiles and ventricular rate profile during AF.
- Remote device follow-up with the Medtronic CareLink Network can dramatically reduce the time and resources required for routine device checks. Patients can use the Medtronic CareLink Service to transmit comprehensive device data using a standard phone line while at home, work or while traveling. Within minutes, clinics can view patient device data from any Internet-enabled PC using the secure clinician website. Learn more.
- Advanced patient alert capabilities, including lead integrity, battery condition change, and ICD therapy issues.
Mechanical Specifications
| Volume |
|
38 cc |
| Size: H x W x D |
|
73 x 51 x 14 mm |
| Mass |
|
77 g |
| Pace/sense ports |
|
Three IS-1 bipolar (A, RV, LV) |
| Defibrillation ports |
|
Two DF-1 (RV coil [HVB], SVC [HVX]) |
| External shield |
|
Titanium |
| Radiopaque ID |
|
PRJ |
| Battery |
|
Lithium silver vanadium oxide; 3.2 V nominal |
Medtronic CareLink® Network
Remote device follow-up with the Medtronic CareLink Network can dramatically reduce the time and resources required for routine device checks. Patients can use the Medtronic CareLink Service to transmit comprehensive device data using a standard phone line while at home, work or while traveling. Within minutes, clinics can view patient device data from any Internet-enabled PC using the secure clinician website. Learn more.
References
- Wathen MS, Sweeney MO, DeGroot PJ, et al. Shock reduction using antitachycardia pacing for spontaneous rapid ventricular tachycardia in patients with coronary artery disease. Circulation. August 14, 2001;104:796-801.
- Accuracy based on Positive Predictive Value (PPV). Wilkoff B, Gillberg J, DeSouza C. The Enhanced PR Logic dual chamber tachyarrhythmia detection algorithm: retrospective analysis of supraventricular tachycardia with long PR intervals. JACC Abstract #$873-4, Feb. 2001.
|