Sprint Quattro™ 6944

Steroid-Eluting, True Bipolar, Passive Fixation ICD Lead

True Bipolar Sensing and Two Defibrillation Coils

  • Quadripolar technology
  • More effective re-detection of VF1

More Efficient Pacing

  • Greater than 750Ω pacing impedance2
  • Low pacing thresholds-median pacing threshold at 1.0 V-0.20 ms
    at 3 months2
  • Steroid elution for reduced inflammation at the tissue interface

The new Sprint Quattro Haloe™ tip electrode, with greater than 750Ω impedance,2 is steroid eluting and platmized

Easy Introduction, Navigation and Handling

  • Small 9 Fr introducer size (without guide wire)
  • Isodiametric—even where lead body meets the coils
  • Small lead body (8.2 Fr)
  • Isoglide™ overlay provides slippery surface

Reliability

  • Patented compression lumens designed to resist crush
  • Built on the solid experience and reliability of over 80,000 Sprint leads sold worldwide

Specifications

Sprint Quattro™ 6944
Polarity Quadripolar
Sensing True bipolar
Fixation Passive (tined)
Lead lengths 58, 65, 75, 100 cm
Connectors 2 DF-1 unipolar high voltage
1 IS-1 bipolar pace/sense
Insulation Silicone with Isoglide™ overlay
Tubing design Multilumen with compression lumens
Tip electrode Platinized steroid-eluting Haloe™ design
Defibrillation electrode shadow areas 483 mm2 RV coil
677 mm2 SVC coil
Electrode surface areas 1.6 mm2 tip
17 mm2 ring
585 mm2 RV coil
819 mm2 SVC coil
Interelectrode spacing 8 mm tip-ring
12 mm tip-RV coil
180 mm tip-SVC coil
Electrode lengths RV coil: 5.7 cm
SVC coil: 8.0 cm
Diameters 8.2 Fr body (2.7 mm)
8.2 Fr (2.7 mm) RV/SVC coils
Introducer 9 Fr without guide wire
10.5 Fr with guide wire
Conductor resistances
(65 cm length)
4.45Ω Cathode coil, nominal
39.6Ω Anode cable, nominal
1.2Ω to RV coil
0.8Ω to SVC coil, maximum
Stylets Purple (0.016 in.)
Gray (0.014 in.)
Serial number prefix TDC
References
  1. Natale A, Sra J, Axtell K, et al. Undetected ventricular fibrillation in transvenous implantable cardioverter-defibrillators: Prospective comparison of different lead system-device combinations. Circulation. 1996;93:91-98.
  2. Medtronic Sprint Quattro™ Model 6944 ICD Lead Clinical Study.