Sprint Quattro 6944
Steroid-Eluting, True Bipolar, Passive Fixation
ICD Lead
True Bipolar Sensing and Two Defibrillation Coils
- Quadripolar technology
- More effective re-detection of VF1
More Efficient Pacing
- Greater than 750Ω pacing impedance2
- Low pacing thresholds-median pacing threshold at 1.0 V-0.20 ms
at 3 months2
- Steroid elution for reduced inflammation at the tissue interface
The new Sprint Quattro Haloe tip electrode, with greater than 750Ω impedance,2 is steroid eluting and platmized
Easy Introduction, Navigation and Handling
- Small 9 Fr introducer size (without guide wire)
- Isodiametriceven where lead body meets the coils
- Small lead body (8.2 Fr)
- Isoglide overlay provides slippery surface
Reliability
- Patented compression lumens designed to resist crush
- Built on the solid experience and reliability of over 80,000 Sprint leads sold worldwide
Specifications
| Sprint Quattro 6944 |
| Polarity |
Quadripolar |
| Sensing |
True bipolar |
| Fixation |
Passive (tined) |
| Lead lengths |
58, 65, 75, 100 cm |
| Connectors |
2 DF-1 unipolar high voltage
1 IS-1 bipolar pace/sense |
| Insulation |
Silicone with Isoglide overlay |
| Tubing design |
Multilumen with compression lumens |
| Tip electrode |
Platinized steroid-eluting Haloe design |
| Defibrillation electrode shadow areas |
483 mm2 RV coil
677 mm2 SVC coil |
| Electrode surface areas |
1.6 mm2 tip
17 mm2 ring
585 mm2 RV coil
819 mm2 SVC coil |
| Interelectrode spacing |
8 mm tip-ring
12 mm tip-RV coil
180 mm tip-SVC coil |
| Electrode lengths |
RV coil: 5.7 cm
SVC coil: 8.0 cm |
| Diameters |
8.2 Fr body (2.7 mm)
8.2 Fr (2.7 mm) RV/SVC coils |
| Introducer |
9 Fr without guide wire
10.5 Fr with guide wire |
Conductor resistances
(65 cm length) |
4.45Ω Cathode coil, nominal
39.6Ω Anode cable, nominal
1.2Ω to RV coil
0.8Ω to SVC coil, maximum |
| Stylets |
Purple (0.016 in.)
Gray (0.014 in.) |
| Serial number prefix |
TDC |