Medtronic
 
 

Tachyarrhythmia
Management

Sprint™ 6945

Specifications

Sprint 6945
Polarity Tripolar
Sensing Integrated bipolar
Fixation Extendable/retractable active (helix)
Lead lengths 58, 65, 75, 100 cm
Connectors 2 DF-1 unipolar high voltage
1 IS-1 bipolar pace/sense
Insulation Silicone
Tubing design Multilumen with compression lumens
Tip electrode CapSureFix® steroid-eluting
Defibrillation electrode shadow areas 491 mm2 RV coil
785 mm2 SVC coil
Electrode surface areas 6.3 mm2 tip
350 mm2 RV coil
568 mm2 SVC coil
Interelectrode spacing Tip-RV coil: 12 mm
Tip-SVC coil: 18 cm
Electrode lengths RV coil: 5 cm
SVC coil: 8 cm
Diameters 7.8 Fr body (2.6 mm)
3.12 mm RV/SVC coils
2.7 mm tip
Introducer 10.5 Fr (without guide wire)
12 Fr (with guide wire)
Conductor resistances
(65 cm length)
30Ω to pacing electrode
1.5Ω to coil

Additional Information