Medtronic
 
 

Tachyarrhythmia
Management

Medtronic Steroid-Eluting ICD Leads

Clinical Results

Sprint Lead 69471 69442 69323 69424 69435 69456
Lead Survival Probability7 not applicable 100%
(>30 days)8
99%
(42 months)
99%
(33 months)
97%
(33 months)
99%
(30 months)
Single-lead implant success 100% (n=80) 100% (n=112) 98% (n=114 of 116) 100% (n=98) 94% (n=61 of 66) 100% (n=78)
Mean DFT 6 J (median) (n=79) 9 J (median) (n=99) 9.4 J (n=94) 8.4 J (n=94) 10.3 J (n=56) 8.2 J (n=67)
Patients with DFTs < 9 J 73%
(n=58 of 79)
73%
(n=72 of 99)
64%
(n=60 of 94)
76%
(n=71 of 94)
54%
(n=30 of 56)
78%
(n=52 of 67)
Patients with DFTs < 15 J 95%
(n=75 of 79)
96%
(n=95 of 99)
92%
(n=86 of 94)
95%
(n=89 of 94)
89%
(n=50 of 56)
96%
(n=64 of 67)
Spontaneous episode efficacy of ventricular origin 99% 99% 96% 100% 100% 100%
Lead-related complications 1% 3% 3% 2% 2% 0%
References
  1. Medtronic Sprint Quattro Secure™ Model 6947 ICD Lead Clinical Study
  2. Medtronic Sprint Quattro™ Model 6944 ICD Lead Clinical Study
  3. Medtronic Sprint 6932 Clinical Study
  4. Medtronic Sprint 6942 Clinical Study
  5. Medtronic Sprint 6943 Clinical Study
  6. Medtronic Sprint 6945 Clinical Study
  7. Medtronic Tachyarrhythmia Product Performance Report. August 1, 2001
  8. Early Sprint Quattro™ 6944 TCSS chronic data (>30 days) show cumulative survival of 100%