Medtronic.com | Physician | Cardiology | Home |
Tachyarrhythmia
Management |
|
|
|
Medtronic Steroid-Eluting ICD Leads
Clinical Results
| Sprint Lead |
69471 |
69442 |
69323 |
69424 |
69435 |
69456 |
| Lead Survival Probability7 |
not applicable |
100%
(>30 days)8 |
99%
(42 months) |
99%
(33 months) |
97%
(33 months) |
99%
(30 months) |
| Single-lead implant success |
100% (n=80) |
100% (n=112) |
98% (n=114 of 116) |
100% (n=98) |
94% (n=61 of 66) |
100% (n=78) |
| Mean DFT |
6 J (median) (n=79) |
9 J (median) (n=99) |
9.4 J (n=94) |
8.4 J (n=94) |
10.3 J (n=56) |
8.2 J (n=67) |
| Patients with DFTs < 9 J |
73%
(n=58 of 79) |
73%
(n=72 of 99) |
64%
(n=60 of 94) |
76%
(n=71 of 94) |
54%
(n=30 of 56) |
78%
(n=52 of 67) |
| Patients with DFTs < 15 J |
95%
(n=75 of 79) |
96%
(n=95 of 99) |
92%
(n=86 of 94) |
95%
(n=89 of 94) |
89%
(n=50 of 56) |
96%
(n=64 of 67) |
| Spontaneous episode efficacy of ventricular origin |
99% |
99% |
96% |
100% |
100% |
100% |
| Lead-related complications |
1% |
3% |
3% |
2% |
2% |
0% |
References
- Medtronic Sprint Quattro Secure Model 6947
ICD Lead Clinical Study
- Medtronic Sprint Quattro Model 6944 ICD Lead
Clinical Study
- Medtronic Sprint 6932 Clinical Study
- Medtronic Sprint 6942 Clinical Study
- Medtronic Sprint 6943 Clinical Study
- Medtronic Sprint 6945 Clinical Study
- Medtronic Tachyarrhythmia Product Performance Report.
August 1, 2001
- Early Sprint Quattro 6944 TCSS chronic data
(>30 days) show cumulative survival of 100%
|
|