Medtronic.com | Physician | Cardiology | Home |
Tachyarrhythmia
Management
|
|
|
|
Medtronic CareLink® Programmer Specifications
Power Requirement
| Voltage |
100 to 127 VAC nominal or 200 to 240 VAC nominal |
| Frequency |
50/60 Hz nominal |
Printer
| Resolution |
600 dots/25 mm horizontally for waveform traces at print speeds through
25 mm/sec,
400 dots/25 mm at 50 mm/sec;
200 dots/25 mm vertically |
| Paper Type |
110 mm wide, Thermal, Medtronic No. 6092 |
| Waveform Recording Speeds |
12.5,25 or 50 mm/s. |
ECG Specifications
| Defibrillation Protection |
IEC 60601-2-27 and AAMI-EC11 |
| Input Impedance at DC |
>2.5 Meg Ohms at 10 Hz |
| ECG Range of Linear Operation |
–5 mV to +5 mV |
| ECG Total Signal Error |
<10% |
| Pacing Pulse Width |
>100 µ to 2 ms |
Temperature Limits
| Operating |
48.9 degrees F to 95 degrees F
(9.4 degrees C to 35 degrees C) |
| Storing |
–7.6 degrees F to 140 degrees F
(–22 degrees C to 60 degrees C) |
Modem
| Command Set |
Hayes and Microcom compatible AT commands |
| Data Interface |
PCMCIA Release 2.1 compliant, 68-pin connector |
| Analog Interface |
RJ-11 modular telephone connector |
| DTE Speeds |
1200, 2400, 4800, 7200, 9600, 14400, 19200, 37400, 57600, 115200 bps
Other |
Screen Size
15 inches
|
|