Healthcare Professionals

Arctic Front Advance

Cardiac CryoAblation Catheter

Indications, Safety, and Warnings

Indications

The Arctic Front Advance™ Cardiac CryoAblation Catheter system is indicated for the treatment of drug refractory recurrent symptomatic paroxysmal atrial fibrillation.

Contraindications

Use of Arctic Front Advance™ cryoballoon is contraindicated 1) In the ventricle because of the danger of catheter entrapment in the chordae tendinae 2) In patients with one or more pulmonary vein stents, 3) In patients with cryoglobulinemia, 4) In patients with active systemic infections and 5) In conditions where the manipulation of the catheter within the heart would be unsafe (e.g. intracardiac mural thrombus).

Warnings/Precautions

Do not re-sterilize this device for purpose of reuse. Use only the 12 Fr FlexCath ® Steerable Sheath family with the Arctic Front Advance cryoballoon because using another sheath may damage the catheter or balloon segment. Do not inflate the balloon inside the sheath. Always verify with fluoroscopy or by using the proximal shaft visual marker that the balloon is fully outside the sheath before inflation to avoid catheter damage. Do not position the cryoballoon catheter within the tubular portion of the pulmonary vein to minimize phrenic nerve injury and pulmonary veins stenosis. Do not connect the cryoballoon to a radiofrequency (RF) generator or use it to deliver RF energy because this may cause catheter malfunction or patient harm. The catheter contains pressurized refrigerant during operation, release of this gas into the circulatory system due to equipment failure or misuse could result in gas embolism, which can occlude vessels and lead to tissue infarction with serious consequences. Always advance and withdraw components slowly to minimize the vacuum created and therefore minimize the risk of air embolism. Do not pull on the catheter, sheath, umbilical cables, or console while the catheter is frozen to the tissue; this may lead to tissue injury. Do not advance the balloon beyond the guide wire to reduce the risk of tissue damage. Do not pass the catheter through a prosthetic heart valve (mechanical or tissue) to avoid damage to the valve, valvular insufficiency or premature failure of the prosthetic valve. Always inflate the balloon in the atrium then position it at the pulmonary vein ostium to avoid vascular injury. Do not ablate in the tubular portion of the pulmonary vein. Use continuous phrenic nerve pacing throughout each cryoablation application in the right pulmonary veins. Placing a hand on the abdomen in the location of the diaphragm to assess for changes in the strength of the diaphragmatic contraction or loss of capture, to avoid nerve injury. In case of no phrenic nerve capture, frequently monitor diaphragmatic movement using fluoroscopy. Stop ablation immediately if phrenic nerve impairment is observed. The Arctic Front Advance cryoballoon was not studied for safety of changes in anticoagulation therapy in patients with paroxysmal atrial fibrillation. This equipment should be used only by or under the supervision of physicians trained in left atrial cryoablation procedures. Cryoablation procedures should be performed only in a fully equipped facility.

Potential Complications

Potential complications/adverse events from cardiac catheterization and ablation include, but are not limited to the following: Anemia; Anxiety; Atrial flutter; Back pain; Bleeding from puncture sites; Blurred vision; Bradycardia; Bronchitis; Bruising; Cardiac tamponade; Cardiopulmonary arrest; Cerebral vascular accident; Chest discomfort/pain/pressure; Cold feeling; Cough; Death; Diarrhea; Dizziness; Esophageal damage (including esophageal fistula); Fatigue; Fever; Headache; Hemoptysis Hypotension/hypertension; Lightheadedness; Myocardial infarction; Nausea/vomiting Nerve injury; Pericardial effusion; Pulmonary vein stenosis; Shivering; Shortness of breath; Sore throat; Tachycardia; Transient ischemic attack; Urinary infection; Vasovagal reaction; Visual changes

Refer to the device manual for detailed information regarding the procedure, indications, contraindications, warnings, precautions, and potential complications/adverse events. For further information, please contact your Medtronic sales rep.

Caution: Swedish law restricts this device to be ordered by, and sold to, a physician or medical institution only.