| General Characteristics |
|---|
| Unit dimensions: |
213 mm. (h) x 137 mm. (w) x 55 mm. (d) (8.385 in. (h) x 5.393 in. (w) x 2.165 in. (d))
(8.385 in. (h) x 5.393 in. (w) x 2.165 in. (d))
|
| Unit weight: |
1 kg |
| |
2.2 lb |
| Power Supply |
|---|
| Input Voltage 100-240 V, 50/60 Hz, ± 10% |
|
| Battery |
|---|
| Battery Type |
Lithium Ion |
| Battery Pack Nominal Voltage and Energy |
7.2V; 18.72Wh |
| Battery Operation |
3 hours for removable bettery pack; 20 minutes for internal battery pack |
| Battery Charging Time |
Up to 5 hours when the monitor is off and up to 8 hours when the monitor is on, for both batteries |
| Display |
|---|
| Screen |
109mm (4.3inch) Color TFT Display
Pixel Pitch: 0.198 (horizontal) x 0.198(vertical) mm
Active Display Area: 95.04 (horizontal) x 53.856 (vertical) mm
Resolution 480 x 272 pixels
Viewing angle (vertical) 125°
Viewing angle (horizontal) 140°
|
| Trace Speed |
3.0, 6.3, 12.5 and 25 mm/sec |
| Waveform Sampling Rate |
77.82 samples/sec for SpO2 (fixed)
20 samples/sec for Capnography (fixed)
|
| Trend Storage |
48 hours with 1 second resolution |
| Trend Display |
Graphical Display: 1h, 2h, 4h, 8h, 12h views
Tabular Display intervals: 1 sec, 5 sec, 15 sec, 30 sec, 1 min, 5 min, 15 min, 30 min, 1 hour
|
| Algorithms |
|---|
| Integrated Pulmonary Index™ algorithm, Nellcor™ SatSeconds Technology, Apnea-Sat Alert™ algorithm, Smart Alarm For Respiratory Analysis™ |
| Communication Interface |
|---|
| Wi-Fi (Vital Sync ™ virtual patient monitoring platform connectivity option) and file transfer via USB flash drive or micro SD card |
| Microstream™ Capnography |
|---|
| CO2 Units |
mmHg, kPa or Vol%. |
| CO2, etCO2 Range |
0-150 mmHg |
| CO2 Waveform Resolution |
0.1 mmHg |
| etCO2 Resolution |
1 mmHg |
| CO2 Accuracy |
0-38 mmHg: ± 2 mmHg
39-150 mmHg: ± (5% of reading + 0.08 for every 1 mmHg above 38 mmHg)
|
| Respiration Rate Range |
0-150 bpm |
| Respiration Rate Accuracy |
0-70 bpm: ±1 bpm
71-120 bpm: ±2 bpm
121-150 bpm: ±3 bpm
|
| CO2 Alarms |
No breath, EtCO2 high, EtCO2 low, RR high, RR low, IPI low
(IPI also requires pulse oximetry information)
|
| Flow Rate |
50 (42.5 ≤ flow ≤ 65) ml/min, flow measured by volume |
| Waveform Sampling |
20 samples/s |
| Response Time |
2.95 s (typical); with use with sampling lines with long tubing, ~5.0 s |
| Initialization Time |
40 s (typical) |
| Calibration Interval |
Initially calibrate after 1,200 operating hours, then once a year or after 4,000 operating hours, whichever comes first |
| Nellcor™ Pulse Oximetry |
|---|
| SpO2 Measurement range |
1-100% |
| Sp02 Accuracy |
|---|
| Adult and Pediatric Modes |
|
| SpO2 range 70%-100% |
±2 digits over the range of 70 to 100% (when using the accessories defined in this document), including under low perfusion; with motion, ±3 digits; with low saturation (60-80%) ±3 digits |
| Infant/Neonatal Mode |
| SpO2 range 70%-100% |
±2% over the range of 70 to 100% (when using the accessories defined in this document), with motion ±3%; with low saturation (60-80%) ±3% |
| Pulse Rate Range |
20 to 250 beats per minute
Pulse Rate values of < 20 beats per minute shall be displayed as 0 beats per minute
Pulse Rate values of > 250 beats per minute shall be displayed as 250 beats per minute.
|
| Pulse Rate Accuracy |
±3 digits over the range of 20 to 250 beats per minute inclusive,
including under low perfusion; with motion, 48 to 127 bpm ±5 digits
|
| Alarms |
SpO2 High, SpO2 Low, PR High, PR Low |
| Sat Sec Range |
10 - 100 |