Efficacy of real-time computer-aided detection of colorectal neoplasia in a randomized trial

Title Efficacy of real-time computer-aided detection of colorectal neoplasia in a randomized trial
First author Alessandro Repici (Department of Gastroenterology, Humanitas Research Hospital, Milano, Italy)
Journal Gastroenterology
Pub date May 2020 (online), August 2020 (print)
DOI 10.1053/j.gastro.2020.04.062


Abstract

Objective

Evaluate the safety and efficacy of computer-assisted detection (CADe) in the real-time identification of colorectal neoplasia during colonoscopy. (Artificial Intelligence for Colorectal Adenoma Detection: AID Study).

Study design

A multicenter, randomized study where patients underwent colonoscopy for colorectal cancer screening, post-polypectomy surveillance, workup following positive fecal immunohistochemical test, or GI symptoms. 700 subjects were randomized in a 1:1 fashion to colonoscopy with and without CADe using GI-Genius™ intelligent endoscopy module (15 patients were excluded from final analysis, N= 685).

Key findings

341 patients were randomized to CADe, 344 to standard colonoscopy. No statistical differences were detected between the two groups for reason for colonoscopy, adequate bowel preparation, or cecal intubation rate. 49.2% of patients were men; mean age 61.32 ± 10.2 years.
 
CADe was 30% more likely to detect adenomas/colorectal cancer than colonoscopy without (RR=1.3, 95% CI 1.14-1.45):
  • 54.8% of colonoscopy patients with CADe were diagnosed (n=187)
  • 40.4% of control patients were diagnosed (n=139)
CADe was 28% more likely to detect proximal (proximal to the splenic flexure) adenomas/colorectal cancer than colonoscopy without:
  • 36.1% CADe compared to 28.2% control (RR=1.28, 95% CI=1.03-1.59)
CADe was more likely to detect both non-polypoid and polypoid lesions than control:
  • 26.6% CADe versus 18.4% control in non-polypoid lesions (RR=1.42, 95% CI=1.09-1.79)
  • 37.3% CADe versus 26.5% control in polypoid lesions (RR=1.36, 95% CI=1.12-1.62)
Sessile serrated lesion detection was not statistically significant between the two groups. Adenomas per colonoscopy (APC) were higher for the CADe group compared to control:
  • 1.07 ± 1.54 average APCs for CADe arm versus 0.71 ± 1.20 average APCs for control arm (Incidence Rate Ratio=1.46, 95% CI=1.15-1.86)
Withdrawal times were the same between the two groups (417± 101 seconds for CADe vs. 435 ± 149 seconds for controls, p=0.1). 27.1% of overall polyp resections were non-neoplastic; the CADe and control arms were not statistically different in identification of non-neoplastic lesions (RR=1.00, 95% CI = 0.90-1.12, p=0.940).

Conclusion

The study revealed that computer-assisted detection (CADe) in the real-time identification of colorectal neoplasia during colonoscopy had an increased adenoma detection rate than the control group. The use of CADe did not impact the withdrawal time, nor was there a change in the detection of non-neoplastic lesions.


More ways to learn about AI and connect

Photo of a 3D illustration of a network abstract background.

Artificial intelligence
in healthcare

Learn about the world of artificial intelligence (AI). Discover Genius Academy.

Photo of a 3D illustration of a network abstract background.

The future is now —
AI in gastroenterology

The benefit of GI Genius™ system introduces new possibilities in the field of endoscopy.

Photo of a 3D illustration of a network abstract background.

Sign up

Subscribe to receive updates on the latest advancements and physician perspectives of AI in healthcare.

US-DG-2100149 ©2021 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. All other brands are trademarks of a Medtronic company.

Disclaimer: All content from healthcare professionals is their individual conclusions, unless otherwise cited. All speaker or author engagement for content is noted to acknowledge funding from Covidien LP, a Medtronic company, for any consulting engagement.