Follow These Steps to Return the Device to the pre-MRI Configuration
Note: The device maintains the parameters that were set while initiating the SureScan feature until the MRI SureScan parameter is programmed to Off after the MRI scan.
After the MRI scan is complete, MRI SureScan must be programmed to Off using the 2090 programmer. This restores the device parameter values to the pre-MRI SureScan configuration.
Perform the following steps to program MRI SureScan to Off:
- Program the MRI SureScan parameter to Off.
- Select [PROGRAM].
- Select [Close] to return to the Additional Features window. Select [OK] to return to the Parameters screen.
The device parameter values are now restored to the pre-MRI SureScan configuration.
Note: During each interrogation, the device is monitored for possible electrical reset conditions and disabled therapies. If a condition is detected that requires attention, the programmer displays a Device Status Indicator warning in a pop-up window and on the Quick Look screen.