Overview

Portable and accurate measurement technology

The Nellcor™ N-85 portable pulse oximetry monitor with Microstream™ technology provides both accurate SpO2 and etCO2 values in a single, easy-to-use, handheld device.

Microcap™ handheld capnographs deliver an easy-to-use assessment of respiratory status, whether intubated or non-intubated, in any clinical setting.

Features

Versatile design for fast, simple setup and use

Handheld monitors that are convenient for spot checks and continuous monitoring in a variety of settings, including EMS/ED, transport.

Order Information

Order Information
Order Code Description Unit of Measure Quantity
N85 Nellcor™ N-85 Hand-Held Capnograph/Pulse Oximeter with OxiMax™ Technology Each 1
Monitors
CS04178 Microcap™ Handheld Capnograph - North America/U.S.  Each 1
CS04179 Microcap™ Handheld Capnograph - Outside North America/Non-U.S.
Each 1
Accessories
CS04321 Carrying Case Each 1
CS07779 Monitor Protective Boot Each 1
CS03743 Clamp/Mounting Bracket Each 1
CS03736 Rechargeable Battery Pack (replacement for original battery pack) Each 1
CS04934 Medical Grade AC Adapter (replacement for original AC adapter) - US Each 1
CS08610 Medical Grade AC Adapter (replacement for original AC adapter) - EU Each 1
CS08505 12 Volt cable Each 1
CS07143 D to A Converter (typically used to interface to sleep labs) Each 1
CS04284 Communication Adapter Kit Each 1
CS08644 MSM Interface Kit (Bernoulli, Oxinet) Each 1
CS08642 Nurse Call Interface Kit Each 1
Specifications
Nellcor™ N-85 Hand-Held Capnograph/Pulse Oximeter with OxiMax™ Technology
Display Range 0-99 mmHg, 0-13.2 kPa and 0-13.0 Vol% at sea level
Sampling flow rate 50 ± 7.5 mL/min
Warm-up time 30 seconds typical, reaches steady-state
Accuracy 0 to 38 mm Hg ±2 mm Hg
  39 to 99 mm Hg ±5% of reading
Interface RS-232 interface at 9600 baud; full duplex serial interface via RJ-45 connector
Modes Adult and Neonate
Alarms Adjustable alarm limits for all parameters; audio level as defined in ISO 9703-2 standard
Display Graphic LCD display (128 x 64 dots) with LED backlight (contrast and backlight control); two numeric fields, three digits each, using 7-segment LED; two-color LED alarm bar
Microcap™ Handheld Capnograph
Sampling flow rate 50 ml/ min
Warm-up time 30 seconds typical
Accuracy 0-38 mmHg ± 2 mmHg
  39-99 mmHg ± 2 mmHg ± (5% of expected reading in mmHg +[0.08 x (expected reading in mmHg –39mmHg)])
  Expected reading = (barometric pressure in mmHg) x (%CO2 (in Vol%)) x 0.97
  Equivalent values for kPa and Vol%
  Respiratory rate 0-150 breaths/min
Modes Adult and Neonate
Alarms Adjustable alarm limits for etCO2, FiCO2, and respiratory rate
Display Graphic LCD (128 x 64 dots) with LED backlight, (contrast and backlight control)
  Bright, seven-segment LED for numeric display of etCO2 and respiratory rate
  Two color alarm bar
Interface Nurse Call RS232 interface between monitor and remote systems
  Bernoulli MSM Nurse Call System easily wired via stereo phone jack
  Philips®* Patient Monitor Interface: Microcap® only - one parameter (v3.15 and higher)