Micro catheters

Marathon™ flow-directed micro catheter

Micro catheters

Marathon™ flow-directed micro catheter

The Marathon™ flow-directed micro catheter is designed for the infusion of therapeutic agents and diagnostic materials during peripheral and neurovascular procedures.

Description

The maneuverability of Marathon™

The Marathon™ flow-directed micro catheter is a single-lumen catheter designed for the injection of embolic agents and diagnostic materials in tortuous, distal vessels. The catheter has a highly flexible distal shaft to facilitate the advancement of the catheter in the anatomy.1

Features

Engineered for control2-4

Robust tensile strength and kink resistant during Onyx™ LES injection, due to nitinol braiding and stainless steel coil design2,5

Smooth catheter handling during navigation and catheter removal due to proximal pushability2,3

Excellent burst profile

Minimum static burst pressure of 150 PSI after one hour of plug-and-push technique4


Compatible with Mirage 0.008" hydrophilic guidewire and Onyx liquid embolic system. Maximum guidewire: 0.010" 2

Ordering information

Item number Stylet Outer diameter (Fr) Inner diameter (in) Usable length (cm) Distal coated length (cm) Maximum guidewire (in)
105-5056 Without 2.7 > 1.5 0.015 > 0.013 165 25 0.010

Similar products



  1. Marathon™ Flow Directed Micro Catheter Instructions for Use: 70876-001 Rev. 03/14.
  2. Based on internal test report TR03-106, Design Verification Marathon AVM Catheter Model 105-5055. Rev A.
  3. Based on internal test report TR03-106ADD2, Marathon AVM Micro Catheter Design Verification Testing Addendum Test Report. Rev A.
  4. Based on internal test report D00229257, Apollo and Marathon Hub Design Change Design Verification Report. Rev A.
  5. Based on internal drawing DWGS105-5056, Marathon165cm AVM Micro Catheter (no stylet) drawing. Rev J.