Brain sensors

BIS™ pediatric sensor

<p>The BIS™ pediatric sensor, intended for children age 4 and above, captures EEG signals from the frontal brain cortex to be displayed on the BIS™ monitor.</p>

Features

The BIS™ pediatric sensor:

  • Is a noninvasive sensor applied to the patient’s forehead
  • Is designed for the smaller heads of pediatric patients
  • Offers peel-and-stick simplicity and Zipprep™ technology
  • Collects electroencephalograph (EEG) data using BIS™ technology

How does our sensor technology work to capture the low-­voltage EEG signal?

  1. We use a conductive ink that is printed directly on the surface of the sensor, under the adhesive foam and Zipprep™ technology. This creates an electrode surface that is helpful in picking up the low­-voltage EEG.
  2. The Zipprep™ technology helps clear away the first layer of the epidermis. The mechanical action of pressing on the electrode results in the tines clearing away some of the first layer of dead skin cells, exposing the inner, more electrically conductive layer of skin. The conductive gel within the electrode permeates into this newly exposed area, creating a good electrical pathway between the EEG-­carrying inner layers of skin and the conductive traces within the sensor. The EEG signal is then carried through these traces to the BIS™ monitor.
  3. We include a thin layer of sponge that contains a very precise amount of gel. This gel creates a bridge between the forehead and the conductive electrode surface.

These combined components help achieve an optimal environment to acquire and maintain the EEG signal.

Specifications

Shelf life One and a half years from date of manufacture
Connector (paddle) Polycarbonate plastic (no patient contact)
Tab stiffener Polycarbonate sheet (no patient contact), one side coated with adhesive
Module Copper, brass, Kapton®* (no patient contact)
Flexible circuit Polyester screen-printed with cured Ag/AgCI and dielectric inks (incidental patient contact)
Electrode gel Potassium chloride (KCI) aqueous gel
Electrode tines disk Green nylon plastic
Electrode sponge disk Polyurethane, polyester foam
Electrode tines retaining disk 1/32 inch white polyethylene foam (no patient contact), both sides coated with medical-grade adhesive
Basepad 1/16 inch white polyethylene foam, both sides coasted with medical-grade adhesive
Liner card Silicone-coated, high-impact, polystyrene (no patient contact)
Packaging Paper, aluminum foil, polyethylene (no patient contact)