Item number | Description | Units per box |
---|---|---|
PM100N | Monitor | 1 |
PM100N-HC | Monitoring system — includues monitor, 8-hour battery, homecare kit and power cord | 1 |
PM100N-MAXN | Monitoring system — includes monitor, 8-hour battery, Nellcor™ SPO2 adhesive sensors and power cord | 1 |
PT00155475 | 16-hour battery | 1 |
Bedside patient monitoring
Nellcor™ bedside SpO₂ patient monitor with home care and sleep study modes
<p>The Nellcor™ bedside SpO₂ patient monitor with home care and sleep study mode provides at-a-glance access to real-time patient respiratory status information.</p>
Features
- Displays plethysmographic waveforms, pulse amplitude, and real-time SpO2 and pulse rate data
- 96-hour trend memory with data captured every four seconds for quick review of changes in patient status
- Includes SatSeconds™ alarm management, a clinician-controlled feature that has been proven to reduce nuisance alarms1
- Password protected home care and sleep study modes
- The device meets IEC 60601-1-11 for home health equipment to promote the safe and effective use of the monitor by lay users in the non-hospital setting
Specifications
Performance — measurement range | |
SpO2 | 1% to 100% |
Pulse rate | 20 to 250 beats per minute (bpm) |
Pulse amplitude | 0.03% to 20% |
Measurement accuracy — saturation | |
Adult | 70% to 100% ± 2 digits |
Adult and neonate low sat | 60% to 80% ± 3 digits |
Neonate | 70% to 100% ± 2 digits |
Low perfusion | 70% to 100% ± 2 digits |
Adult and neonate with motion | 70% to 100% ± 3 digits |
Pulse rate | |
Adult and neonate | 20 to 250 bpm ± 3 digits |
Low perfusion | 20 to 250 bpm ± 3 digits |
Adult and neonate with motion | 20 to 250 bpm ± 5 digits |
Electrical — instrument | |
Power requirements | 100 to 240 VAC, 50/60 Hz, 45 VA |
Fuse rating | Fast-acting 2A 32VAC/DC, Fast-acting 500mA 32VAC/50DC |
Electrical — battery | |
Type | Lithium ion |
Battery capacity | Minimum of eight hours using new, fully charged battery with no alarms; optional 16-hour battery available. |
Environmental — temperature | |
Operating conditions | 5 C to 40 C (41 F to 104 F) |
Transport and storage (in shipping carton) | -20 C to 60 C (-4 F to 140 F) |
Environmental — humidity | |
Operating conditions | 15% to 93% noncondesending |
Transport and storage (in shipping carton) | 15% to 93% noncondesending |
Environmental — altitude | |
Operating conditions | -170 m to 4,877 m (-557 ft to 16,000 ft) |
Transport and storage (in shipping carton) | -304 m to 6,096 m (-1,000 ft to 20,000 ft) |
Physical characteristics | |
Weight | 1.6 kg (3.5 lbs including battery) |
Size | 82 H × 255 W × 155 D (mm), (3.23 H ×10.04 W × 6.10 D (in)) |
Equipment classifications | |
Type of protection against electric shock | Class 2 (internally powered) |
Degree of protection against electric shock | Type BF — applied part |
Mode of operation | Continuous |
Liquid ingress | IP 22 |
Output | |
Trend data download via wired or USB for archiving or data analysis | |
Display/indicators | |
Pulse amplitude indicator (eight segments) | |
Visual indicators | Pulse search, audible alarms silenced or off, interference indicator, battery charging, and SatSeconds alarm management clock, pleth wave form |
Alarms | |
SatSeconds alarm management | |
Audible and visual alarms for high/low saturation and pulse rate, low battery, sensor off, and sensor disconnect | |
Categories | Patient status and system status |
Priorities | Low, medium and high |
Notification | Audible and visual |
Setting | Default, institutional, and last setting |
Alarm system delay | < 10 seconds |
Optional accessories | |
|
|
Available modes | |
|
|
Connectivity | |
|
|
Simple setup and maintenance | |
The Nellcor™ bedside SpO2 patient monitoring system is RoHs compliant, and enables hospital staff to set institutional defaults, replace the battery, perform diagnostics to troubleshoot performance issues, and perform on‑site maintenance on the monitor. |
Ordering information
Similar products
-
Nellcor™ bedside SpO₂ patient monitorThe Nellcor™ bedside SpO₂ patient monitor offers a compact design, improved functionality, and an easy-to-read color user interface.
-
Nellcor™ bedside respiratory patient monitoring system
The Nellcor™ bedside respiratory patient monitoring system provides easy access to the status of a patient’s oxygenation and respiration.
The Nellcor™ pulse oximetry monitoring system should not be used as the sole basis for diagnosis or therapy and is intended only as an adjunct in patient assessment.
- Brostowicz HM, Khodayar Rais-Bahrami K. Oxygen saturation monitoring in the neonatal intensive care unit: evaluation of a new alarm management. In: Program and abstracts of the American Academy of Pediatrics National Conference & Exhibition; October 17–20, 2009; Washington, D.C.