Item number | Description | Units per box |
---|---|---|
PM10N-AD3 | PM10N-AD3 PM10N KIT WITH 3YR WARRANTY | 1 |
PMAC10N-B | PMAC10N-B BLU PROTECT COVER PORTABLSPO2 | 1 |
PMAC10N-G | PMAC10N-G GRN PROTECT COVER PORTABLSPO2 | 1 |
PMAC10N-N | PMAC10N-N NVY PROTECT COVER PORTABLSPO2 | 1 |
PMAC10N-P | PMAC10N-P PNK PROTECT COVER PORTABLSPO2 | 1 |
PMAC10N-T | PMAC10N-T TRANSPORT COVER PORTABL SPO2X1 | 1 |
Portable patient monitoring
Nellcor™ portable SpO₂ patient monitor
<p>The Nellcor™ portable SpO₂ patient monitor is a convenient, handheld pulse oximeter ideal for spot-checks or continuous monitoring.</p>
Features
- Small, lightweight, ergonomic handheld monitor for continuous and spot-check pulse oximetry monitoring
- Supports a robust data set, including real-time SpO2 and pulse rate values, SatSeconds alarm management, pleth waveform, blip bar, and tabular trend data
- Supports sleep study mode, which enables the clinician to dim the LCD display and silence alarms to prevent disruption of the patients’ sleep
- Compatible with the complete line of Nellcor™ single-patient use, reusable and specialty sensors with OxiMax™ technology
- 80 hours of data storage capability
- Supports data export to an external personal computer for data analysis and printing functions
- Supports standard and home-care modes for use in hospital, hospital-type facilities, transport, mobile environments, and home-care environments
Note: Transport and mobile environments are defined as intra hospital transport or transport between facilities or between a facility and a home environment
Clinical applications
Providing accurate, reliable SpO2 and pulse rate readings, this portable, easy-to-use device is an ideal solution for these critical screenings.
Specifications
Enclosure |
|
Weight | 274 g (0.604 lbs), including four batteries |
Dimensions | 70 mm W × 156 mm H × 32 mm D (2.76 in W × 6.14 in H × 1.26 in D) |
Display | |
Screen size | 88.9 mm (3.5 in), measured diagonally |
Screen type | TFT LCD, white LED backlight, viewing cone of 60 degree and optimal viewing distance of 1 meter |
Resolution | 320 × 480 pixels |
Alarms | |
Categories | Patient status and system status |
Priorities | Low, medium, and high |
Notification | Audible and visual |
Setting | Default, institutional, and last setting |
Alarm volume level | 45–80 db |
Alarm system delay | <10s |
Range and accuracy |
|
Measurement ranges | |
SpO2 saturation range | 1% to 100% |
Pulse rate range | 20 to 250 beats per minute (bpm) |
Perfusion range | 0.03% to 20% |
Display sweep speed | 6.25 mm/sec |
Measurement accuracy |
|
Saturation | |
Adult | 70% to 100% ±2 digits |
Adult and neonate low sat | 60 to 80% ±3 digits |
Neonate | 70 to 100% ±2 digits |
Low perfusion | 70 to 100% ±2 digits |
Adult and neonate with motion | 70 to 100% ±3 digits |
Pulse rate | |
Adult and neonate | 20 to 250 bpm ±3 digits |
Low perfusion | 20 to 250 bpm ±3 digits |
Adult and neonate with motion | 20 to 250 bpm ±5 digits |
Electrical |
|
Battery | Four new lithium batteries with 3,000 mAh will typically provide 20 hours of monitoring with no external communication, no audible alarm sound, and at an ambient temperature of 25 C. |
Type | Lithium AA |
Voltage | 1.5V × 4 |
Equipment classifications |
|
---|---|
Type of protection against electric shock | Class I (internally powered) |
Degree of protection against electric shock | Type BF — applied part |
Mode of operation | Continuous |
Electromagnetic compatibility | IEC 60601-1-2:2007 IEC 60601-1-2:2014 |
Ingress protection | IP22: Protected against foreign objects and moisture |
Degree of safety | Not suitable for use in the presence of flammable anesthetics |
Ordering information
Related links
The Nellcor™ pulse oximetry monitoring system should not be used as the sole basis for diagnosis or therapy and is used only as an adjunct in patient assessment.
This product is licenced with Health Canada as "NELLCOR PORTABLE SPO2 PATIENT MONITORING SYSTEM" on Health Canada licence number "94759".