No Image Available

Ventilator filters and HMEs

DAR™ HME

<p><sup></sup>The DAR™&nbsp;HME minimizes heat and moisture loss from patient airways during anesthesia or respiratory care.</p>

Features

  • ISO standard 15 mm and 22 mm fittings
  • Available with end tidal CO2 port, which offers convenient access to airway gases
  • Available with integrated extendable connector to simplify setup

Specifications

Item Number 353/19004 353S19007 353P5908 353S19046
Item description TRACHEOLIFE II HYGROLIFE II FLEXLIFE HME TRACHEOLIFE III
Tidal volume ≥ 100 ml 100 - 1000 ml. 200 - 1000 ml. ≥ 100 ml
Moisture loss (9360-1) 11 mg H2O/L at Vt 500 mL 8 mgH2O/L. at Vt. 500 ml. 5 mgH2O/L. at Vt. 500 ml. 11 mg H2O/L at Vt 500 mL
Resistance to flow before use (ISO 9360-1) (ISO 23328-2)
30 L per minute 0.9 cmH2O 1,0 cmH2O 0,4 cmH2O 0.6 cm H2O
60 L per minute 2.0 cmH2O 2,3 cmH2O 1,1 cmH2O 1.1 cm H2O
90 L per minute 3.1 cmH2O 3,8 cmH2O 2,0 cmH2O 1.7 cm H2O

Internal volume

(ISO 9360-1)

(ISO 23328-2)

13 ml. 29 ml. 76 - 91 ml. 13 ml.
Weight 9 gr. 22 gr. 12 gr. 8 gr.

Ordering information

DAR™ HME filters

Item number Description Conn 1 Conn 2 Eas/Box
332/5115 DAR tm FIXED CATHETER MT 15F 15M 25
332/5663 DAR tm EXTEND FIXED CATHETER 15F 15M 25
610/5257 DAR tm double-swivel COBB CONNECTOR 22M/15M 15M 10
331/5661 DAR tm double EXTEND SWIVEL CATHETER 22M/15M 15M 25

DAR™ HME filter accessories

Item number Description Connection 1 Connection 2 Units per box
332/5115 DAR™ fixed catheter 15F 15M 25
332/5663 DAR™ extendible fixed catheter 15F 15M 25
610/5257 DAR™ double-swivel connector 22M/15M 15M 10
331/5661 DAR™ double-swivel connector, extendible 22M/15M 15M 25

† Internal testing Mirandola (various 2005–2008). All technical information is sourced from information for use (IFU) documents PT00144742, PT00191236, PT00162624, and clinical evaluation report CER10055277. Please reference the IFU for complete product details.