Overview

Innovative solutions for fragile lungs

The Puritan Bennett™ NeoMode 2.0 software option helps clinicians provide safe ventilatory support to neonates weighing as little as 300 grams.

The NeoMode 2.0 software was developed expressly for neonates to address the issues most critical to their care and safety, such as accurate breath delivery, responsive triggering, lower elevated oxygen preset for procedures, effective alarm management, and automatic leak compensation.

Managing Neonatal Stressors

Working together to reduce infant stress in the NICU. Because today decides tomorrow.

Learn More

Features

Solutions and settings for small lungs

The Puritan Bennett™ NeoMode 2.0 software option offers a full spectrum of functionality that helps clinicians effectively ventilate neonatal patients.

This software option includes features that allow clinicians to:

  • Set a tidal volume as small as 2 mL for neonatal patients weighing as little as 300 grams
  • Adjust the ideal body weight (IBW) setting without disconnecting the patient or interrupting ventilation
  • Exercise greater control over the FiO2 setting by temporarily increasing FiO2 by only 20%
  • Deliver effective noninvasive ventilation, including the use of SIMV and CPAP
  • Adjust tidal volume in increments as small as 0.1 cc, when ventilating with volumes between 2 mL and 5 mL

The NeoMode 2.0 software option also enables responsive breath triggering by allowing clinicians to adjust the flow sensitivity from 0.1 to 10 L/min.

Specifications
Expiratory sensitivity (ESENS)
Range: 1% to 80%
New patient value: 25%
Flow pattern
Range: square or descending ramp
New patient value: descending ramp
IBW
Range: 0.3 to 7.0 kg (0.66 to 15.4 lbs)
New patient value: 3.0 kg (6.6 lbs)
Resolution: 0.1 kg (0.2 lbs) for 0.3 to 3.5 kg (0.66 to 7.7 lbs), 0.5 kg (1.1 lbs) for 3.5 to 7 kg (7.7 to 15.4 lbs)
Inspiratory time (TI)
Range: 0.20 to 8.00s
New patient value: based on VT and VMAX
Resolution: 0.01s
Mandatory breath type
Range: volume control (VC) or pressure control (PC)
New patient value: PC
Peak flow (VMAX)
Range: 1.0 to 30 L/min
New patient value: based on IBW
Resolution: 0.1 for 1 to 20 L/min; 1 for 20 L/min and above
Respiratory rate (f)
Range: 1.0 to 150 breaths per minute
New patient value: 20 breaths per minute
Resolution: 0.1 for 1.0 to 10 breaths per minute; 1 for 10 to 150 breaths per minute
Spontaneous breath type
Range: pressure supported (PS) or not pressure supported (none)
New patient value: PS
Tidal volume (VT)
Range: 2 to 315 mL
Resolution: 0.1 for 2 to 5 mL
  1 for 5 to 100 mL
  5 for 100 to 400 mL
New patient value: the greater of 5 mL or 7.25 lbs x IBW
Trigger type  
Range: flow (VTRIG) only
New patient value: VTRIG
Interface type
Neonatal nasal CPAP prongs
Endotracheal tube
Tracheostomy tube
Compliance and approvals
The NeoMode 2.0 software option complies with the requirements of these standards: IEC 60601-1 (EN 60601-1), IEC 60601-2-12, EN 794-1, UL 2601-1, CSA C22.2 M601.1 and EN 1441