Aortic stent grafts

Endurant™ II and Endurant™ IIs stent graft system

<p>Endurant stent graft system is designed to treat abdominal aortic aneurysms (AAA) using a minimally invasive endovascular aneurysm repair (EVAR).</p>


Features

The Endurant™ system was designed with key features3 to address sac regression and ensure durability.4–6

Accurate placement and controlled deployment4

  • Intuitive graft deployment system7 offers controlled delivery at the intended target zone with 99.3% delivery and deployment success investigational device exemption (IDE).†,4

  • Tip capture deployment mechanism — even after deployment of three stent rings — provides gradual, controlled, and accurate deployment at the intended sealing zone.4


Deployment of the Endurant™ II stent graft


Deployment of the Endurant™ IIs stent graft



Continuous seal5

  1. Separation of seal from fixation — with suprarenal fixation, which enables seal to occur independently5
  2. M-shaped proximal stents, unique to the Endurant™ stent graft, maximize wall apposition and circumferential conformability5 to minimize in-folding resulting in low type Ia endoleak rates2
  3. Electropolished nitinol stents maximize circumferential conformability with dynamic continuous seal5

The Endurant™ stent graft features a continuous seal.


Resistance against type II endoleak6

  • Woven, high-density, multifilament polyester material8 that promotes sac thrombosis6
  • Addresses sac regression3–6
The Endurant™ stent graft offers durability in aortic procedures.


Decade of durability data2​

Endurant™ AAA stent graft system: The first and only EVAR system with a decade of global registry outcomes.2



The Endurant™ family of products

Three configurations

  • Bifurcated main body stent grafts:
    • Endurant™ II bifurcated configuration — an aortoiliac stent graft that is available in three lengths.
    • Endurant™ IIs bifurcated configuration — an aortic configuration available in a single, shorter length. 
  • Aorto-uni-iliac (AUI) main body stent graft:
    • The AUI device is indicated for the endovascular treatment of infrarenal abdominal aortic or aortoiliac aneurysms only in patients whose anatomy does not allow the use of a bifurcated device. The proximal end of the AUI configuration is deployed into the proximal neck and upper section of the aneurysm. The proximal section of the AUI configuration is composed of nitinol stents sewn to a fabric graft.
Endurant™ II stent graft is designed to treat abdominal aortic aneurysms (AAA) in both challenging and straightforward patient anatomy.

Endurant™ II stent graft   

Endurant™ IIs stent graft is a three-piece system that is designed to treat abdominal aortic aneurysms (AAA) in both challenging and straightforward anatomy.

Endurant™ IIs stent graft   

The Endurant™ II stent graft system aorto-uni-iliac (AUI) stent graft is indicated for the endovascular treatment of infrarenal abdominal aortic or aortoiliac aneurysms in patients whose anatomy does not allow the use of a bifurcated stent graft.

Endurant™ II AUI stent graft   

Ordering information

Endurant™ II bifurcations

Item number Proximal graft diameter (mm) Distal graft diameter (mm) Distal design Total covered length (mm) Delivery system Catheter outer diameter (Fr)
ETBF2313C124EE 23 13 C 124 EE 18
ETBF2313C145EE 23 13 C 145 EE 18
ETBF2313C166EE 23 13 C 166 EE 18
ETBF2316C124EE 23 16 C 124 EE 18
ETBF2316C145EE 23 16 C 145 EE 18
ETBF2316C166EE 23 16 C 166 EE 18
ETBF2513C124EE 25 13 C 124 EE 18
ETBF2513C145EE 25 13 C 145 EE 18
ETBF2513C166EE 25 13 C 166 EE 18
ETBF2516C124EE 25 16 C 124 EE 18
ETBF2516C145EE 25 16 C 145 EE 18
ETBF2516C166EE 25 16 C 166 EE 18
ETBF2813C124EE 28 13 C 124 EE 18
ETBF2813C145EE 28 13 C 145 EE 18
ETBF2813C166EE 28 13 C 166 EE 18
ETBF2816C124EE 28 16 C 124 EE 18
ETBF2816C145EE 28 16 C 145 EE 18
ETBF2816C166EE 28 16 C 166 EE 18
ETBF2820C124EE 28 20 C 124 EE 18
ETBF2820C145EE 28 20 C 145 EE 18
ETBF2820C166EE 28 20 C 166 EE 18
ETBF3216C124EE 32 16 C 124 EE 20
ETBF3216C145EE 32 16 C 145 EE 20
ETBF3216C166EE 32 16 C 166 EE 20
ETBF3220C124EE 32 20 C 124 EE 20
ETBF3220C145EE 32 20 C 145 EE 20
ETBF3220C166EE 32 20 C 166 EE 20
ETBF3616C145EE 36 16 C 145 EE 20
ETBF3616C166EE 36 16 C 166 EE 20
ETBF3620C145EE 36 20 C 145 EE 20
ETBF3620C166EE 36 20 C 166 EE 20

Endurant™ IIs bifurcations

Item number Proximal graft diameter (mm) Distal graft diameter (mm) Distal design Total covered length (mm) Delivery system Catheter outer diameter (Fr)
ESBF2314C103EE 23 14 C 103 EE 18
ESBF2514C103EE 25 14 C 103 EE 18
ESBF2814C103EE 28 14 C 103 EE 18
ESBF3214C103EE 32 14 C 103 EE 20
ESBF3614C103EE 36 14 C 103 EE 20

Limbs

Item number Distal graft diameter (mm) Distal graft diameter (mm) Distal design Total covered length (mm) Delivery system Catheter outer diameter (Fr) Total contralateral covered length with EII/EIIs bifurcated§ Total ipsilateral covered length with EIIs bifurcated
ETLW1610C82EE 16 10 C 82 EE 14 136 155
ETLW1610C93EE 16 10 C 93 EE 14 147 166
ETLW1610C124EE 16 10 C 124 EE 14 178 177–197
ETLW1610C156EE 16 10 C 156 EE 16 210 209–229
ETLW1610C199EE 16 10 C 199 EE 16 253 252–272
ETLW1613C82EE 16 13 C 82 EE 14 136 155
ETLW1613C93EE 16 13 C 93 EE 14 147 166
ETLW1613C124EE 16 13 C 124 EE 14 178 177–197
ETLW1613C156EE 16 13 C 156 EE 16 210 209–229
ETLW1613C199EE 16 13 C 199 EE 16 253 252–272
ETLW1616C82EE 16 16 C 82 EE 14 136 135–155
ETLW1616C93EE 16 16 C 93 EE 14 147 146–166
ETLW1616C124EE 16 16 C 124 EE 14 178 177–197
ETLW1616C156EE 16 16 C 156 EE 16 210 209–229
ETLW1616C199EE 16 16 C 199 EE 16 253 252–272
ETLW1620C82EE 16 20 C 82 EE 16 136 155
ETLW1620C93EE 16 20 C 93 EE 16 147 166
ETLW1620C124EE 16 20 C 124 EE 16 178 177–197
ETLW1620C156EE 16 20 C 156 EE 16 210 209–229
ETLW1620C199EE 16 20 C 199 EE 16 253 252–272
ETLW1624C82EE 16 24 C 82 EE 16 136 155
ETLW1624C93EE 16 24 C 93 EE 16 147 166
ETLW1624C124EE 16 24 C 124 EE 16 178 177–197
ETLW1624C156EE 16 24 C 156 EE 16 210 209–229
ETLW1624C199EE 16 24 C 199 EE 16 253 252–272
ETLW1628C82EE 16 28 C 82 EE 16 136 155
ETLW1628C93EE 16 28 C 93 EE 16 147 166
ETLW1628C124EE 16 28 C 124 EE 16 178 177–197
ETLW1628C156EE 16 28 C 156 EE 16 210 209–229
ETLW1628C199EE 16 28 C 199 EE 16 253 252–272

‡ The limb mates with the AUI stent graft on the ipsilateral side.

§ These calculations assume the minimum 30 mm overlap between the bifurcated stent graft and the contralateral iliac limb per the Endurant™ stent graft system instructions for use.

◊ The 3–5 stent overlap is available only with select limbs. Please refer to the instructions for use for more information.

Aortic extensions

Item number Proximal graft diameter (mm) Distal graft diameter (mm) Distal design Total covered length (mm) Delivery system Catheter outer diameter (Fr)
ETCF2323C49EE 23 23 C 49 EE 18
ETCF2525C49EE 25 25 C 49 EE 18
ETCF2828C49EE 28 28 C 49 EE 18
ETCF3232C49EE 32 32 C 49 EE 20
ETCF3636C49EE 36 36 C 49 EE 20

Abdominal tubes

Item number Proximal graft diameter (mm) Distal graft diameter (mm) Distal design Total covered length (mm) Delivery system Catheter outer diameter (Fr)
ETTF2323C70EE 23 23 C 70 EE 18
ETTF2525C70EE 25 25 C 70 EE 18
ETTF2828C70EE 28 28 C 70 EE 18
ETTF3232C70EE 32 32 C 70 EE 20
ETTF3636C70EE 36 36 C 70 EE 20

AUI

Item number Proximal graft diameter (mm) Distal graft diameter (mm) Distal design Total covered length (mm) Delivery system Catheter outer diameter (Fr)
ETUF2314C102EE 23 14 C 102 EE 18
ETUF2514C102EE 25 14 C 102 EE 18
ETUF2814C102EE 28 14 C 102 EE 18
ETUF3214C102EE 32 14 C 102 EE 20
ETUF3614C102EE 36 14 C 102 EE 20

Iliac extensions

Item number Proximal graft diameter (mm) Distal graft diameter (mm) Distal design Total covered length (mm) Delivery system Catheter outer diameter (Fr)
ETEW1010C82EE 10 10 C 82 EE 14
ETEW1313C82EE 13 13 C 82 EE 14
ETEW2020C82EE 20 20 C 82 EE 16
ETEW2424C82EE 4 24 C 82 EE 16
ETEW2828C82EE 8 28 C 82 EE 18

† Delivery success = successful vascular access and delivery of the device to the intended anatomical treatment site. Deployment success = successful deployment of the device in the intended site and the successful removal of the delivery system.

  1. Based on internal Endurant™ patient implant tracking at Medtronic. Data current as of September 2023.
  2. Lalys F, et al. J Vasc Surg. 2017;65(6):1830–1838.
  3. Makaroun M, et al. J Vasc Surg. 2011;54:601–608.
  4. Based on internal test report DVR7245 Design Verification Report: Endurant Stent Graft System conducted January–July 2007.
  5. Wakefield T, et al. J Vasc Surg. 1989;9:234–243.
  6. Verhagen, et al. The ENGAGE Registry: Ten-Year Outcomes with the Endurant Stent Graft for Endovascular Abdominal Aortic Aneurysm Repair. Presented at: Charing Cross 2023 International Symposium; April 26, 2023; London, UK.

  1. Rouwet EV, et al. Eur J Vasc Endovasc Surg. 2011;42:489–497
  2. Based on internal materials report 10011927DOC Rev.AE.