Ventilator filters and HMEs

DAR™ HME

<p><sup></sup>The DAR™&nbsp;HME minimizes heat and moisture loss from patient airways during anesthesia or respiratory care.</p>

Features

  • ISO standard 15 mm and 22 mm fittings
  • Available with end tidal CO2 port, which offers convenient access to airway gases
  • Available with integrated extendable connector to simplify setup

Specifications

  Adult–pediatric small — Hygrolife II Adult–pediatric foam HME with integrated catheter mount — flexlife
Tidal volume 100–1000 mL 200–1000 mL
Moisture loss (9360-1) 8 mgH20/L at Vt 500 mL 5 mgH20/L at Vt 500 mL
Resistance to flow before use (ISO 9360-1) (ISO 23328-2)
30 L per minute 1.0 cmH2O 0.4 cmH2O
60 L per minute 2.3 cmH2O 1.1 cmH2O
90 L per minute 3.8 cmH2O 2.0 cmH2O

Internal volume

(ISO 9360-1)

(ISO 23328-2)

29 mL 76— 91 mL
Weight 22 g 12 g

Ordering information

DAR™ HME filters

Item number Size Patient end connection Machine end connection Units per box
353U5413 Adult–pediatric, small 22M/15F 22F/15M 50
353U5908 Adult–pediatric foam HME with integrated catheter mount 15F 22F 25

DAR™ HME filter accessories

Item number Description Connection 1 Connection 2 Units per box
332U5115 DAR™ extension tube 15F 15M 25
332U5663 DAR™ extendible extension tube 15F 15M 25
610U5257 DAR™ double-swivel connector 22M/15M 15M 50
331U5661 DAR™ double-swivel connector, extendible 22M/15M 15M 25

Similar products

All technical information is sourced from various information for use (IFU) documents and clinical evaluation reports. Please reference the instructions for use (IFU) for a complete product details.

† Internal testing Mirandola (various 2005–2008). All technical information is sourced from information for use (IFU) documents PT00144742, PT00191236, PT00162624, and clinical evaluation report CER10055277. Please reference the IFU for complete product details.