Ventilator filters and HMEs

DAR™ mechanical filter HME

<p>The DAR™ mechanical filter HME helps protect against cross-contamination while minimizing heat and moisture loss from patient airways.</p>

Features

  • Hygroscopic membrane traps patient’s moisture providing effective airway humidification
  • End tidal carbon dioxide (CO2) sampling port offers convenient access to airway gases
  • ISO standard 15 mm and 22 mm fitting connects with breathing system

Specifications

  Adult large — Hygroster
Tidal volume 300—1500 mL
Moisture loss (ISO 9360-1) 5 mgH20/L at Vt 500 mL
Resistance to flow before use (ISO 9360-1) (ISO 23328-2)
30 L per minute 1.2 cmH2O
60 L per minute 2.6 cmH2O
90 L per minute 4.2 cmH2O
Filtration efficiency
NaCl (ISO 23328-1) 99.7%
Bacterial Greater than or equal to 99.9999%
Viral Greater than or equal to 99.9999%

Internal volume

(ISO 9360-1)

(ISO 23328-2)

95 mL
Weight (approximate) 49 g

Ordering information

DAR™ mechanical filters HME

Item number Size Patient end connection Machine end connection Units per box
354U5876 Pediatric—adult 22M/15F 22F/15M 50

DAR™ mechanical filter HME accessories

Item number Description Filter end Machine end Units per box
332U5115 DAR™ extension tube 15F 15M 25
332U5663 DAR™ extendible extension tube 15F 15M 25
610U5257 DAR™ double-swivel connector 22M/15M 15M 50
331U5661 DAR™ double-swivel connector, extendible 22M/15M 15M 25

Similar products

All technical information is sourced from various information for use (IFU) documents and clinical evaluation reports. Please reference the instructions for use (IFU) for a complete product details.

† Internal testing Mirandola (various 2005–2008). All technical information is sourced from information for use (IFU) documents PT00141155, PT00162624, and clinical evaluation report CER10055277. Please reference the IFU for complete product details.