Healthcare Professionals

NC Sprinter RX

Noncompliant Balloon Dilatation Catheter

NC Sprinter RX Noncompliant Balloon Dilatation Catheter

The NC Sprinter® RX Noncompliant Rapid Exchange Balloon Dilatation Catheter combines tried and true Sprinter technology with the high-pressure capability and controlled expansion of a noncompliant balloon.

How the NC Sprinter RX Works 

Stent implantation often follows balloon angioplasty. Once the stent is implanted, post stent deployment as assessed by IVUS or fluoroscopy may be completed to ensure adequate stent expansion and apposition against the arterial wall.1

Post-dilatation of a stent with a noncompliant balloon

Features

The NC Sprinter RX noncompliant rapid exchange balloon dilatation catheter combines tried and true Sprinter technology with the high-pressure capacity and controlled expansion of a noncompliant balloon.2

NC Sprinter RX features 3.0 mm NC Sprinter RX Compliance

Engineered for Exact Sizing to Enhance Stent Apposition

Engineered for exact sizing

Expand All

Specifications

Specifications

Catheter length 138 cm
Lesion entry profile (in.) 0.016
Tip length (mm)
  • 2.00 – 4.00
  • 4.50 – 5.00

3.5 mm
4.5 mm
Balloon material Soft Fulcrum Plus
Coating Selective DuraTrac
Gold-swaged marker bands Double
Nominal pressure 10 atm
Rated burst pressure 18 atm
MiniWrap folds
  • 2.00–3.75 mm
  • 4.00–5.00 mm

3 folds
5 folds
Shaft dimensions for 2.0 – 3.75 mm
  • Proximal
  • Distal

1.9 F
2.4 F / 2.6 F
Shaft dimensions for 4.00 – 5.00 mm
  • Proximal
  • Distal

1.9 F
3.0 F
MGCID (in.)
  • 0.056 (2.00 – 4.00 mm)
  • 0.066 (4.50 – 5.00 mm)

Diameters (mm)

2.00 – 5.00
Lengths (mm) 6, 9, 12, 15, 21, 27

 

Crossing Profile

Balloon Diameter (mm) Crossing Profile OD (in)
2.00 0.024
2.25 0.024
2.50 0.025
2.75 0.025
3.00 0.026
3.25 0.027
3.50 0.027
3.75 0.028
4.00 0.029
4.50 0.030
5.00 0.032

Compliance Table

Pressure (atm) Balloon Diameter (mm)
2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00 4.50 5.00
8 1.98 2.17 2.45 2.70 2.97 3.09 3.36 3.62 3.87 4.27 4.75
9 2.00 2.20 2.49 2.74 3.01 3.15 3.42 3.68 3.93 4.34 4.84
10 2.02 2.23 2.51 2.77 3.04 3.18 3.45 3.72 3.97 4.39 4.91
11 2.03 2.24 2.53 2.80 3.07 3.22 3.49 3.76 4.01 4.44 4.97
12 2.05 2.26 2.55 2.82 3.10 3.25 3.52 3.79 4.05 4.48 5.01
13 2.07 2.28 2.57 2.84 3.13 3.28 3.55 3.82 4.08 4.51 5.06
14 2.08 2.30 2.58 2.87 3.15 3.31 3.58 3.85 4.11 4.55 5.10
15 2.10 2.32 2.60 2.89 3.18 3.33 3.61 3.88 4.15 4.58 5.14
16 2.12 2.33 2.62 2.91 3.20 3.36 3.63 3.90 4.18 4.61 5.18
17 2.13 2.36 2.63 2.93 3.23 3.39 3.65 3.93 4.22 4.65 5.22
18 2.15 2.38 2.65 2.95 3.26 3.41 3.68 3.96 4.25 4.68 5.26
19 2.17 2.40 2.66 2.98 3.29 3.44 3.71 3.99 4.29 4.72 5.30
20 2.19 2.42 2.68 3.00 - 3.47 3.73 4.02 - 4.76 -
21 2.21 2.44 2.70 3.03 - 3.50 3.77 4.05 - 4.80 -
22 2.23 2.46 2.72 3.05 - 3.53 - 4.08 - 4.84 -
Nominal pressure
Rated burst pressure

Test data on file with Medtronic, Inc.
* Rated burst pressure: the maximum pressure to which a balloon is designed to be inflated (95% confidence that 99.9% will not fail below rated burst pressure). DO NOT EXCEED.
Nominal pressure: the pressure at which the balloon reaches its labeled diameter.

Ordering Information

References

1

Brodie BR. Adjunctive balloon postdilatation after stent deployment: is it still necessary with drug-eluting stents? J Interv Cardiol. 2006;19:43-50.

2

NC Sprinter balloon models 2.25 mm to 5.0 mm are indicated for the post-delivery expansion of balloon expandable stents.

3

Test results based on distal balloon shoulder. Test data on file at Medtronic, Inc. Bench test results may not be indicative of clinical performance.

4

Test data on file at Medtronic, Inc. Bench test results may not be indicative of clinical performance.