NC Stormer OTW

Noncompliant Over-the-Wire Balloon Dilatation Catheter

NC Stormer OTW Noncompliant Over-the-Wire Balloon Dilatation Catheter

When precision and performance are paramount, the NC Stormer® OTW Noncompliant Over-the-Wire Balloon Dilatation Catheter offers controlled compliance, high pressures, comprehensive lengths, and low longitudinal growth.

How the NC Stormer OTW Works

Stent implantation often follows balloon angioplasty. Once the stent is implanted, post stent deployment as assessed by IVUS or fluoroscopy may be completed to ensure adequate stent expansion and apposition against the arterial wall.1

Post-dilatation of a stent with a noncompliant balloon NC Stormer OTW features

Features

When precision and performance are paramount, NC Stormer OTW noncompliant balloon catheter offers:2

  • Controlled compliance
  • High pressures
  • Comprehensive lengths
  • Low longitudinal growth

Low Longitudinal Growth

cv-ncstrmrotw-chrt-low

Test data on file at Medtronic, Inc. Bench test data may not be indicative of clinical performance.

Eight Lengths for Post-dilatation

cv-ncsstormerotw-lengths2

Expand All

Specifications

Specifications

Catheter length 138 cm
Crossing profile (in.)* 0.026
Nominal pressure 2.0 – 4.0 mm 10 atm
Rated burst pressure 2.0 – 4.0 mm 18 atm
Distal / mid-shaft O.D. Distal = 2.5 F
Mid = 2.7 F
Proximal = 3.2 F
Exit markers Brachial 88 cm
Femoral 98 cm
Marker bands Gold swaged
2 on all 2.0 – 4.0 mm O.D.s
Coating Selective DuraTrac coating
No. of balloon folds
2.0 – 2.5 mm
2.75 – 3.0 mm
3.25 – 3.5 mm
3.75 – 4.0 mm


3 folds
4 folds
5 folds
6 folds

Diameters (mm) 2.00 – 4.00
Lengths (mm) 6, 11, 14, 16, 18, 21, 26, 31

* Based on 3.0 mm NC Stormer data. Test data on file at Medtronic. Bench test data may not be indicative of clinical results.

Compliance

  Balloon Diameter (mm)
Pressure (atm) 2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00
4 1.9 2.1 2.3 2.5 2.8 3.0 3.2 3.4 3.5
5 2.0 2.2 2.3 2.6 2.9 3.0 3.3 3.5 3.6
6 2.0 2.2 2.4 2.6 2.9 3.1 3.4 3.5 3.7
7 2.0 2.3 2.4 2.7 3.0 3.1 3.4 3.6 3.8
8 2.0 2.3 2.5 2.7 3.0 3.2 3.5 3.6 3.9
9 2.1 2.3 2.5 2.7 3.0 3.2 3.5 3.7 3.9
10 2.1 2.3 2.5 2.8 3.1 3.3 3.5 3.7 4.0
11 2.1 2.4 2.5 2.8 3.1 3.3 3.5 3.8 4.1
12 2.1 2.4 2.6 2.8 3.1 3.3 3.6 3.8 4.1
13 2.1 2.4 2.6 2.8 3.2 3.3 3.6 3.8 4.1
14 2.2 2.4 2.6 2.8 3.2 3.4 3.6 3.9 4.2
15 2.2 2.4 2.6 2.9 3.2 3.4 3.7 3.9 4.2
16 2.2 2.5 2.6 2.9 3.2 3.4 3.7 3.9 4.2
17 2.2 2.5 2.6 2.9 3.3 3.5 3.7 4.0 4.3
18 2.2 2.5 2.7 3.0 3.3 3.5 3.7 4.0 4.3
19 2.3 2.5 2.7 3.0 3.3 3.5 3.8 4.0 4.4
Nominal pressure
Rated burst pressure*

*Rated burst pressure: the maximum pressure to which a balloon is designed to be inflated (95% confidence that 99.9% will not fail below rated burst pressure). DO NOT EXCEED.
Nominal pressure: the pressure at which the balloon reaches its labeled diameter.

Ordering Information

References

1

Brodie BR. Adjunctive balloon postdilatation after stent deployment: is it still necessary with drug-eluting stents? J Interv Cardiol. 2006;19:43-50.

2

NC Stormer balloon dilatation catheters are indicated for the post-delivery expansion of balloon expandable stents.