Overview

Maximize Viral and Bacterial Filtration for the Puritan Bennett™ 840 Ventilator

The Puritan Bennett™ D/X800 disposable expiratory filter and drain system is designed for the Puritan Bennett™ 840 ventilator. It reduces particles and bacteria in a patient’s exhaled gas and protects the ventilator’s exhalation and spirometry systems.

Product Features

  • Comprehensive condensate drainage system includes one-piece filter and collector vial and optional drain bag
  • Optional drain bag allows for condensation removal without breaking the ventilator circuit
  • Viral and bacterial filtration efficiency of greater than 99.999%
  • Resistance to flow less than 2.5 cmH2O at 100 L/min when new
  • Minimum of 99.97% filtration efficiency of 0.3 microns nominal particle size at 100 L/min flow
  • Meets CDC requirement for n100 filtration equivalency

How to Order

For additional information or to place an order, call Customer Service at +44 (0) 19 2320 2504.

Order Information
Order Code Description Unit of Measure Quantity
4-076887-00 Puritan Bennett™ D/X800 Expiratory Filter Case 12
4-048491-00 Puritan Bennett™ Drain Bag, Single Patient Use Case 25
4-048492-00 Puritan Bennett™ Drain Bag Clamp, Reusable Case 5
4-048493-00 Puritan Bennett™ Drain Bag Tubing, Single Patient Use Case 10
4-074613-00 Puritan Bennett™ Drain Port Cap, Reusable Case 1
Specifications
Filter
Viral/bacterial filtration efficiency Greater than 99.999%
Particle filtration efficiency Minimum of 99.97% of 0.3-μm
  nominal particle size at 100 L/min flow
Inlet Connector 22-mm ISO connector
Resistance to flow Less than 2.5 cmH2O at 100 L/min when new
Flow leakage Less than 0.01 L/min at 140 cmH2O internal pressure
Size 11.7 cm diameter x 12.7 cm long (4.6 in. diameter x 5 in. long)
Internal Volume Approximately 400 mL
Collector Vial
Capacity Approximately 200 mL usable volume (to maximum fill line)
Tubing
Length 17.8 cm (7 in.)
Internal Diameter 0.48 cm (3/16 in.)
Drain Bag
Capacity Approximately 1 L

Order Information

Connect with Medtronic Patient Monitoring & Respiratory Interventions
Your platform for clinical & product educational content. 

 

Medtronic announced that it has decided to keep the Patient Monitoring and Respiratory Intervention businesses as a part of Medtronic, given changing market conditions. These businesses will get increased investment and be combined and called Acute Care & Monitoring (ACM).

Medtronic also made the decision to exit the ventilator product line.

Amid evolving market conditions and a shift to lower acuity ventilators, this decision aligns with Medtronic’s strategy to concentrate resources on our market-leading positions and accelerate innovation-driven growth.

While this decision will take time and planning, we are committed to supporting our existing ventilation customers. We will continue to provide service and support throughout the ventilator life cycle, along with meeting our service contract obligations.
For more information, please visit our website.